Siltberg H, Larsson G, Victor A
Department of Obstetrics and Gynecology, Akademiska Hospital and Uppsala University, Sweden.
Acta Obstet Gynecol Scand. 1998 Nov;77(10):1000-7.
Presently available urodynamic methods are of little use for assessing the severity of urinary stress incontinence or for evaluating of treatment, Cough-induced leak-point pressure may prove to be a more useful urodynamic method in these respects. A vaginal anti-incontinence device was used to validate this new urodynamic method.
The primary aim was to determine the extent to which cough-induced leak-point pressure was affected by the vaginal device. Secondary aims were to study the short-term effects of the vaginal device on leakage and urinary flow, and to relate the change in cough-induced leak-point pressure to the change in leakage as expressed by a short-term pad test.
In a prospective study of 22 women with a history of stress incontinence, the effect of a new vaginal anti-incontinence device (Conveen Continence Guard) on cough-induced leak-point pressure, a short-term pad test and urinary flow was studied.
Cough-induced leak-point pressure increased from 99.9 to 138.9 cm H2O, while leakage, measured by a short-term pad test, decreased from 22.7 to 3.3 g when using the device. Urinary flow was not significantly reduced by the device.
Cough-induced leak-point pressure is a valid, quantitative, dynamic measure of urethral closure function which can be used to study the effects of treatment in patients with stress incontinence. The vaginal device is effective for treatment of stress incontinence in the short-term, without reducing urinary flow.
目前可用的尿动力学方法在评估尿失禁严重程度或评估治疗效果方面作用不大。咳嗽诱发漏尿点压力在这些方面可能是一种更有用的尿动力学方法。使用一种阴道抗失禁装置来验证这种新的尿动力学方法。
主要目的是确定阴道装置对咳嗽诱发漏尿点压力的影响程度。次要目的是研究阴道装置对漏尿和尿流的短期影响,并将咳嗽诱发漏尿点压力的变化与短期护垫试验所显示的漏尿变化相关联。
在一项对22名有压力性尿失禁病史女性的前瞻性研究中,研究了一种新型阴道抗失禁装置(Conveen Continence Guard)对咳嗽诱发漏尿点压力、短期护垫试验和尿流的影响。
使用该装置时,咳嗽诱发漏尿点压力从99.9厘米水柱增加到138.9厘米水柱,而通过短期护垫试验测量的漏尿从22.7克减少到3.3克。该装置未使尿流显著减少。
咳嗽诱发漏尿点压力是一种有效的、定量的、动态的尿道闭合功能测量方法,可用于研究压力性尿失禁患者的治疗效果。阴道装置在短期内对压力性尿失禁有效,且不减少尿流。