Tian Y, Li L C
Hospital Products Division, Abbott Laboratories, Abbott Park, Illinois 60064-3500, USA.
Drug Dev Ind Pharm. 1998 Mar;24(3):275-80. doi: 10.3109/03639049809085620.
The use of a light-scattering particle size distribution analyzer has been shown to be a convenient method for characterizing the particle size distribution of parenteral emulsions. However, the concentrations of the samples used were found to have a major impact on the particle size distribution results, particularly for samples with a mean particle size smaller than 0.2 micron. An increase in sample concentration caused a shift to smaller particle sizes as a result of multiple scattering. The blue-light (tungsten lamp) intensity, instead of the He-Ne laser integrity, should be used to control sample concentration within the optimal range for measurement.
使用光散射粒度分布分析仪已被证明是表征肠胃外乳剂粒度分布的一种便捷方法。然而,发现所用样品的浓度对粒度分布结果有重大影响,特别是对于平均粒径小于0.2微米的样品。由于多次散射,样品浓度的增加导致粒径向更小尺寸偏移。应使用蓝光(钨灯)强度而非氦氖激光完整性来将样品浓度控制在测量的最佳范围内。