Høgh Petersen H, Chen X, Pietersen A, Svendsen J H, Haunsø S
Medical Department B, National University Hospital, 2100 Copenhagen O, Denmark.
Circulation. 1999 Jan 19;99(2):319-25. doi: 10.1161/01.cir.99.2.319.
It is important to increase lesion size to improve the success rate for radiofrequency ablation of ischemic ventricular tachycardia. This study of radiofrequency ablation, with adjustment of power to approach a preset target temperature, ie, temperature-controlled ablation, explores the effect of catheter-tip length, ablation site, and convective cooling on lesion dimensions.
In vitro strips of porcine left ventricular myocardium during different levels of convective cooling and in vivo pig hearts at 2 or 3 left ventricular sites were ablated with 2- to 12-mm-tip catheters. We found increased lesion volume for increased catheter-tip length </=8 mm in vitro (P<0.05) and 6 mm in vivo (P<0. 0001), but no further increase was found for longer tips. For the 4- to 10-mm catheter tips, we found smaller lesion volume in low-flow areas (apex) than in high-flow areas (free wall and septum) (P<0.05). Increasing convective cooling of the catheter tip in vitro increased lesion volume (P<0.0005) for the 4- and 8-mm tips but not for the 12-mm tip as the generator reached maximum output. In contrast to power-controlled ablation, we found a negative correlation between tip temperature reached and lesion volume for applications in which maximum generator output was not achieved (P<0. 0001), whereas delivered power and lesion volume correlated positively (P<0.0001).
Lesion size differs in different left ventricular target sites, which is probably related to convective cooling, as illustrated in vitro. Longer electrode tips increase lesion size for tip lengths </=6 to 8 mm. For temperature-controlled ablation, the tip temperature achieved is a poor predictor of lesion size.
增大病灶大小对于提高缺血性室性心动过速射频消融的成功率很重要。本项关于射频消融的研究,通过调整功率以接近预设目标温度,即温控消融,探讨了导管尖端长度、消融部位和对流冷却对病灶尺寸的影响。
使用尖端长度为2至12毫米的导管,对处于不同对流冷却水平的猪左心室心肌体外条带以及猪活体心脏的2或3个左心室部位进行消融。我们发现,体外情况下,导管尖端长度增加至≤8毫米时病灶体积增大(P<0.05),体内情况下增加至6毫米时病灶体积增大(P<0.0001),但尖端更长时未发现进一步增大。对于4至10毫米的导管尖端,我们发现低血流区域(心尖)的病灶体积比高血流区域(游离壁和室间隔)小(P<0.05)。体外增加导管尖端的对流冷却,对于4毫米和8毫米的尖端,病灶体积增大(P<0.0005),但对于12毫米的尖端,由于发生器达到最大输出,病灶体积未增大。与功率控制消融相反,在未达到发生器最大输出的应用中,我们发现达到的尖端温度与病灶体积呈负相关(P<0.0001),而输送的功率与病灶体积呈正相关(P<0.0001)。
不同左心室目标部位的病灶大小不同,这可能与体外所示的对流冷却有关。电极尖端长度增加至≤6至8毫米时病灶大小增大。对于温控消融,达到的尖端温度对病灶大小的预测性较差。