The effect of overnight wear of graded rigid gas-permeable (RGP) contact lenses on the rabbit corneal epithelial barrier function was assessed by fluorophotometry.
RGP lenses with optimal fitting were worn by rabbits for 24 h. The values of oxygen transmissibility (Dk/L) of lenses used were 0, 10, 27, 43, 84, and 109 x 10(-9) (cm/s)(ml O2/ml mm Hg); lens thickness (0.15 mm) and diameter (14.0 mm) were standardized. Corneal epithelial barrier function was measured by a slit-lamp fluorophotometer. Corneal surface lesions by fluorescein staining were observed by a slit-lamp biomicroscope with the aid of fluorescein filter paper and were scored. Corneal epithelial structure was observed by scanning electron microscopy (SEM).
After 24-h wear of RGP lenses, fluorescein uptake by the cornea increased, and this increase depended on the Dk/L value of lenses worn; mean +/- standard error, 1,601+/-454 (Dk/L 0), 347+/-226 (Dk/L 10), 144+/-42 (Dk/L 27), 100+/-25 (Dk/L 43), 56+/-18 (Dk/L 84), 51+/-15 (Dk/L 109), and 49+/-4 ng/ml (control). There was a significant correlation between the fluorophotometric value and the biomicroscopic scores of corneal lesions. SEM showed progressive desquamation of epithelial superficial and wing cells for lower-value Dk/L lens groups.
This study demonstrated for the first time that overnight wear of Dk/L 84 and 109 lenses does not alter the barrier function of the rabbit corneal epithelium. In contrast, wear of a PMMA lens greatly alters the corneal epithelial barrier function. It is suggested that RGP lenses with a Dk/L >80 can provide safe overnight wear, based on the epithelial barrier function.
通过荧光光度法评估渐变硬性透气性(RGP)隐形眼镜过夜佩戴对兔角膜上皮屏障功能的影响。
将适配良好的RGP镜片给兔佩戴24小时。所用镜片的氧气透过率(Dk/L)值分别为0、10、27、43、84和109×10⁻⁹(厘米/秒)(毫升氧气/毫升毫米汞柱);镜片厚度(0.15毫米)和直径(14.0毫米)标准化。用裂隙灯荧光光度计测量角膜上皮屏障功能。借助荧光素滤纸,用裂隙灯生物显微镜观察荧光素染色的角膜表面损伤并评分。通过扫描电子显微镜(SEM)观察角膜上皮结构。
RGP镜片佩戴24小时后,角膜荧光素摄取增加,且这种增加取决于所佩戴镜片的Dk/L值;平均值±标准误,1601±454(Dk/L 0)、347±226(Dk/L 10)、144±42(Dk/L 27)、100±25(Dk/L 43)、56±18(Dk/L 84)、51±15(Dk/L 109)和49±4纳克/毫升(对照组)。荧光光度值与角膜损伤的生物显微镜评分之间存在显著相关性。SEM显示,较低Dk/L值镜片组的上皮表层和翼状细胞有进行性剥脱。
本研究首次证明,Dk/L值为84和109的镜片过夜佩戴不会改变兔角膜上皮的屏障功能。相比之下,佩戴聚甲基丙烯酸甲酯(PMMA)镜片会极大地改变角膜上皮屏障功能。基于上皮屏障功能,建议Dk/L>80的RGP镜片可安全过夜佩戴。