[Practical use of lithium borate in thermoluminescent dosimetry (author's transl)].

作者信息

Chavaudra J, Marinello G, Brulé A M, Nguyen J

出版信息

J Radiol Electrol Med Nucl. 1976 May;57(5):435-45.

PMID:994091
Abstract

The authors recall the functional principles of thermoluminescent dosimeters: heating, apparatus for measuring the emitted light, circulation of nitrogen, reference source. They take this opportunity to stress the essential role played by the circulation of nitrogen over the dosimeters which equilibrates the temperature of the photo multiplier, reduces the emission of unwanted light, prevents the combustion of dust or other possible impurities and finally improves the reproductibility of the measurements even for high dosod for finding the optimum working conditions for the heating apparatus of the planchette in the most simple T.L.D. readers and in those where the heating apparatus of the planchette has a pre-heating phase. They then study the dosimetric properties of lithium borate incorporated in thin teflon discs (type DLB. 0.13 and 0.4). This shows itself to be very interesting for certain uses because it is a solid dosimeter which does not require annealing between two measurements. The reproductibility of the measurements obtained with this material, the stability of its response relative to the delay between radiation and reading (fading), its response relative to the absorbed dose plus the nature and the energy of the rays, are presented with the usual reservations made for this type of dosimetry. The authors conclude by citing a few applications where they have been able to achieve a satisfactory result with the aid of lithium borate.

摘要

相似文献

1
[Practical use of lithium borate in thermoluminescent dosimetry (author's transl)].
J Radiol Electrol Med Nucl. 1976 May;57(5):435-45.
2
Skin dose with cones and diaphragms in 137Cs teletherapy - a study using lithium borate thermoluminescent dosimeters.
Strahlentherapie. 1975 Aug;150(2):134-7.
3
Energy response of lithium borate thermoluminescent dosimeters.
Health Phys. 1972 Apr;22(4):399-401.
4
[Influence of regenerating and evaluation procedures on the supralinear behavior of LiF thermoluminescent dosimeters].
Strahlenther Onkol. 1988 Apr;164(4):223-7.
5
[The value of routine in vivo measurements in radiotherapy (author's transl)].
J Radiol Electrol Med Nucl. 1977 Nov;58(11):701-8.
6
The suitability of different preparations of thermoluminescent lithium borate for medical dosimetry.
Phys Med Biol. 1982 Aug;27(8):1023-34. doi: 10.1088/0031-9155/27/8/004.
7
LiF and CaF2: Dy thermoluminescent dosimeters.
Strahlentherapie. 1980;156(10):708-13.
8
A TLD system based on lithium borate for the measurement of doses to patients undergoing medical irradiation.
Phys Med Biol. 1976 Jan;21(1):39-51. doi: 10.1088/0031-9155/21/1/003.
9
[The use of thermoluminescent dosimeters for measuring low-energy x-ray and gamma radiation].
Gig Sanit. 1991 Nov(11):50-3.
10
[Use of lithium fluoride in clinical dosimetry. Application to the measurement of the dose delivered to axilla by iridium 192 with endocurietherapy for carcinoma of the breast].
J Radiol Electrol Med Nucl. 1975 Nov;56(11):791-6.