Butter C, Auricchio A, Schwarz T, Fleck E, Schubert B, Hsu W, Michel U, Neuzner J
Virchow Clinic of the Humboldt University, Berlin, Germany.
Pacing Clin Electrophysiol. 1999 Jan;22(1 Pt 2):169-73. doi: 10.1111/j.1540-8159.1999.tb00326.x.
Dual chamber ICD systems use two separate leads for sensing. We developed and tested a new prototype of a single pass dual chamber passive fixation lead for dual chamber ICDs.
The prototype was a modification of the Guidant CPI Endotak DSP lead. The additional sensing electrode for the right atrium consisted of a side-mounted porous atrial ring electrode (AR). Atrial signals were recorded from the lead in patients during normal sinus rhythm (NSR), atrial fibrillation (AFib), and/or atrial flutter (AFl) with the AR in stable contact with the atrial wall or floating. During NSR, with the AR in contact with the atrial wall, an average P wave amplitude of 7.2 +/- 1.5 mV (mean +/- SD, n = 12) was measured. After induction of Afib/AFl, the single amplitude decreased to 3.6 +/- 1.5 mV (n = 8) during AFib and 3.4 +/- 1.7 mV (n = 9) during AFl. Amplitudes dropped between 53% and 75% when the AR lost atrial wall contact. The atrial pacing threshold was 1.0 +/- 0.4 V (n = 16) when the AR was in contact with the atrial wall.
In future dual chamber ICDs the signals from a passive fixation single pass lead could be used for atrial sensing and pacing as long as the sensing electrode for the right atrium remains in contact with the atrial wall. This system might lead to a simpler, less invasive implantation of dual chamber ICD systems.
双腔植入式心律转复除颤器(ICD)系统使用两根独立的导线进行感知。我们研发并测试了一种用于双腔ICD的新型单通道双腔被动固定导线原型。
该原型是对Guidant CPI Endotak DSP导线的改进。右心房的额外感知电极由侧面安装的多孔心房环电极(AR)组成。在正常窦性心律(NSR)、心房颤动(AFib)和/或心房扑动(AFl)期间,当AR与心房壁稳定接触或漂浮时,从患者的导线上记录心房信号。在NSR期间,当AR与心房壁接触时,测得平均P波振幅为7.2±1.5 mV(均值±标准差,n = 12)。诱发Afib/AFl后,AFib期间单振幅降至3.6±1.5 mV(n = 8),AFl期间降至3.4±1.7 mV(n = 9)。当AR失去与心房壁的接触时,振幅下降了53%至75%。当AR与心房壁接触时,心房起搏阈值为1.0±0.4 V(n = 16)。
在未来的双腔ICD中,只要右心房的感知电极与心房壁保持接触,被动固定单通道导线的信号就可用于心房感知和起搏。该系统可能会使双腔ICD系统的植入更简单、侵入性更小。