van der Laarse R, Bruinvis I A, Nooman M F
Acta Radiol Oncol Radiat Phys Biol. 1978;17(2):113-24. doi: 10.3109/02841867809127912.
This report describes now a set of applicators, convering fields with dimensions of 4 to 20 cm, for the 6 to 20 MeV electron beams of a MEL SL75-20 linear accelerator was developed. The electron scatter contribution of the applicator walls to the treatment field was investigated, varying the applicator entrance opening and the scattering foil, with the aim of optimizing the resulting field flatness, with a minimum loss of depth dose. Experiments with field defining end frames and additional perspex scatterers for large field sizes are also reported.
本报告描述了如何为MEL SL75 - 20直线加速器的6至20 MeV电子束开发了一组尺寸为4至20 cm的照射器。研究了照射器壁对治疗野的电子散射贡献,通过改变照射器入口开口和散射箔,目的是在深度剂量损失最小的情况下优化所得野平坦度。还报告了使用场界定端框和用于大野尺寸的额外有机玻璃散射体的实验。