Estimating mean glandular dose using proprietary mammography phantoms.

作者信息

Hartley L D, Cobb B J, Hutchinson D E

机构信息

Radiation Health Section, Health Department of Western Australia, Nedlands, Australia.

出版信息

Appl Radiat Isot. 1999 Jan;50(1):205-13. doi: 10.1016/s0969-8043(98)00025-6.

Abstract

Screening mammography programs prescribe limits on the radiation dose to the glandular tissue of a compressed breast--the Mean Glandular Dose (MGD)--which may vary between 2 and 3 mGy for a 50% adipose and 50% glandular tissue compressed breast of 4.2 to 5.0 cm thickness. Commercially available breast phantoms offer a choice of equivalent tissue thicknesses and composition ranging typically from 30% adipose and 70% glandular tissue to 50% adipose and 50% glandular tissue. A number of these phantom were compared under clinical conditions and the MGD calculated for each using the factors developed by Wu et al. [Wu, X., Barnes, G.T. and Tucker, D.M. (1991) Spectral dependence of glandular tissue dose in screen film mammography, Radiology 179, 143-148].