Digital fundus imaging: a quality and cost comparison with 35-mm film.

作者信息

Prasad S, Bannon P, Clearkin L G, Phillips R P

机构信息

Department of Ophthalmology, Arrowe Park Hospital, Upton, Wirral, UK.

出版信息

Acta Ophthalmol Scand. 1999 Feb;77(1):79-82. doi: 10.1034/j.1600-0420.1999.770118.x.

Abstract

PURPOSE

This study compares 35-mm transparencies with colour digital fundus images.

METHODS

Colour fundus photos were acquired simultaneously on a digital system and on 35-mm film on 32 consecutive patients. These were assessed independently by two ophthalmologists who graded overall picture quality and granularity of the images. They also assessed if the images were adequate for the purpose of making a diagnosis. An analysis of the costs of the two systems was performed.

RESULTS

The film images had significantly better scores for overall picture quality (p=0.0001, 95% CI 0.5 to 1.5) and granularity (p<0.0001, 95% CI 1.5 to 2). However, both digital and film images were considered adequate for diagnosis (p=0.052, 95% CI 0.12 to 8.32).

CONCLUSION

The quality of digital images is acceptable for diagnostic purposes although resolution is limited. Other advantages and developing technologies will make it increasingly attractive in the future.