Inoue Y, Ohtake T, Yokoyama I, Yoshikawa K, Ohtomo K
Department of Radiology, University of Tokyo, Japan.
Nucl Med Commun. 1999 Mar;20(3):267-72. doi: 10.1097/00006231-199903000-00010.
Fractional renal accumulation of a radiotracer is often used to measure renal function with camera-based methods. We evaluated the effect of deadtime count loss on the calculation of fractional renal accumulation of 99Tcm-mercaptoacetyltriglycine (99Tcm-MAG3). In 15 patients, dynamic renal scintigraphy was performed after the injection of 250 MBq of 99Tcm-MAG3. A reference source was placed on the periphery of the field of view of the gamma camera to monitor the loss of counts in dynamic imaging. The activity in the syringe was also counted with the camera to measure the injected dose, and the deadtime count loss in the syringe was computed from the observed count rate and deadtime in imaging a point source. Fractional renal accumulation, the ratio of renal accumulation on dynamic imaging to the injected dose, was assessed with and without correction for deadtime loss. Fractional renal accumulation was over-estimated by 4.1% with no deadtime correction, when the value after correction for loss in both syringe counts and dynamic imaging was used as a standard. It was underestimated by 4.0% with correction for syringe counts only. The results suggest that correction for count loss only in measuring the injected dose does not lead to a significant improvement in the accuracy of estimating renal function with camera-based methods.
放射性示踪剂的肾部分摄取分数通常用于通过基于相机的方法测量肾功能。我们评估了死时间计数损失对99锝-巯基乙酰三甘氨酸(99Tcm-MAG3)肾部分摄取分数计算的影响。对15例患者注射250MBq的99Tcm-MAG3后进行动态肾闪烁显像。在γ相机视野的周边放置一个参考源,以监测动态成像中的计数损失。还用相机对注射器中的放射性活度进行计数以测量注射剂量,并根据观察到的计数率和在对一个点源成像时的死时间计算注射器中的死时间计数损失。在不校正和校正死时间损失的情况下评估肾部分摄取分数,即动态成像时肾摄取与注射剂量的比值。当将注射器计数和动态成像中的损失校正后的数值作为标准时,不进行死时间校正时肾部分摄取分数高估了4.1%。仅校正注射器计数时低估了4.0%。结果表明,仅在测量注射剂量时校正计数损失并不能显著提高基于相机的方法估算肾功能的准确性。