Graham J M, Ashcroft P
Cochlear Implant Programme at Royal National Throat Nose & Ear Hospital, London, England.
Am J Otol. 1999 Mar;20(2):205-8.
Perilymph/cerebrospinal fluid (CSF) "gushers" may occur at cochleostomy during cochlear implant surgery, particularly in patients with congenital cochlear duct malformation in which CSF in the internal auditory meatus is in direct communication with the perilymphatic space in the cochlea. The object of the study was to measure the pressure and flow of a CSF gusher at cochleostomy.
The design was a preoperative pressure measurement.
The setting was a multidisciplinary cochlear implant program.
A 4-year-old girl with bilateral Mondini deformity undergoing cochlear implantation was studied.
A size 23 FG intravenous cannula was inserted into the cochlea and connected to a pediatric drip set to form an improvised manometer.
Intracochlear fluid pressure was measured at 14 cm H2O, equivalent to the normal CSF pressure that would be recorded in a child of this age at lumbar puncture. An indirect measurement of the likely size of the CSF/perilymph defect was made.
This technique may allow better assessment of the risk of postoperative CSF leakage and meningitis.
This simple technique of measuring the pressure in a perilymph gusher can be used to assess the need for careful sealing of the cochleostomy, to measure the reduction in pressure produced by head elevation or a spinal drain, and to assess the probable size of a defect in the lamina cribrosa.
在人工耳蜗植入手术的蜗窗造口术中可能会出现外淋巴/脑脊液“喷射”现象,尤其是在先天性耳蜗管畸形患者中,此类患者内耳道内的脑脊液与耳蜗内的外淋巴间隙直接相通。本研究的目的是测量蜗窗造口术中脑脊液喷射的压力和流量。
设计为术前压力测量。
研究地点为一个多学科人工耳蜗植入项目。
对一名4岁双侧Mondini畸形并接受人工耳蜗植入的女孩进行了研究。
将一根23FG的静脉套管插入耳蜗,并连接到一个儿科输液装置上,以形成一个简易压力计。
测量耳蜗内液体压力为14 cmH₂O,相当于该年龄儿童腰椎穿刺时记录的正常脑脊液压力。对脑脊液/外淋巴缺损的可能大小进行了间接测量。
该技术可能有助于更好地评估术后脑脊液漏和脑膜炎的风险。
这种测量外淋巴喷射压力的简单技术可用于评估蜗窗造口术仔细密封的必要性,测量头部抬高或脊髓引流所产生的压力降低情况,并评估筛板缺损的可能大小。