Five cases of Smeloff-Cutter prosthesis implanted in the mitral position nine years ago were investigated with combined echocardiographic and phonocardiographic recordings. In two cases a malfunction was demonstrated and a surgical confirmation was obtained at the reoperation. Normal and abnormal echocardiographic patterns of the Smeloff-Cutter prosthesis are reported and discussed. The Authors stress the importance of combined echo-phono recordings when malfunction of the mobile element of a mechanical prosthesis is suspected.

摘要

对9年前植入二尖瓣位置的5例斯梅洛夫-卡特人工瓣膜进行了超声心动图和心音图联合记录检查。2例显示有功能障碍,并在再次手术时得到手术证实。报告并讨论了斯梅洛夫-卡特人工瓣膜正常和异常的超声心动图表现。作者强调,当怀疑机械人工瓣膜活动部件出现功能障碍时,联合超声心动图和心音图记录的重要性。