A new platinum balloon-expandable stent (Angiostent) mounted on a high pressure balloon: acute and late results in an atherogenic swine model.

作者信息

Hijazi Z M, Homoud M, Aronovitz M J, Smith J J, Faller G T

机构信息

Department of Pediatrics, Tufts University School of Medicine, Boston, Massachusetts, USA.

出版信息

J Invasive Cardiol. 1995 Jun;7(5):127-34.

Abstract

BACKGROUND

Randomized studies have proven the efficacy and safety of stent placement to treat de novo coronary stenosis. However, the poor radio-opacity and the use of an additional high-pressure balloon to fully expand the stent are the major limitations of the currently clinically-approved stents.

OBJECTIVE

We evaluated the safety, efficacy, angiographic and histologic effect of a new platinum balloon expandable stent mounted on a high-pressure balloon in Yucatan miniature swine fed high cholesterol diet.

METHODS

Fifteen Angiostents (NuMED, Inc., Hopkinton, NY and Angiodynamics, Glens Falls, NY) (coronary stent was 3, 3.5, or 4 mm in diameter and 12 mm long; renal and carotid stents were 5 mm in diameter and 13 mm long) mounted on a high-pressure balloon were placed percutaneously in blood vessels of 10 pigs [5 in circumflex (CX), 2 in left anterior descending (LAD), 5 in renal and 3 in carotid arteries]. The stent was 10-20% larger than the native vessel diameter. All animals received 5000 I.U. of heparin during the procedure and were maintained on 325 mg aspirin daily. Follow-up angiography and histology in the animals was performed at 2, 4, 12, 20, 26 and 52 weeks.

RESULTS

The stents were easily visualized with fluoroscopy and placed in all animals without episodes of balloon rupture or embolization. There was no episode of acute thrombosis. Follow-up angiography in the animals revealed patency of all renal and carotid stents, however, 2/7 coronary stents in the animals revealed angiographic lumen narrowing (> 20%) at 20 and 52 weeks. Histologic examination revealed neointimal formation at the stent site with an average neointimal thickness ranging from 325-650 microns.

CONCLUSION

This stent was safe in this animal model, easily deployed, had excellent radio-opacity and with good short-term patency without anticoagulation. Clinical trials and experience is underway.

摘要

背景

随机研究已证实支架置入术治疗初发冠状动脉狭窄的有效性和安全性。然而,目前临床批准的支架存在不透X线性差以及需要使用额外的高压球囊来充分扩张支架等主要局限性。

目的

我们评估了一种安装在高压球囊上的新型铂合金球囊可扩张支架在喂食高胆固醇饮食的尤卡坦小型猪中的安全性、有效性、血管造影和组织学效果。

方法

将15个血管支架(NuMED公司,纽约州霍普金顿市和Angiodynamics公司,纽约州格伦斯福尔斯市)(冠状动脉支架直径为3、3.5或4毫米,长12毫米;肾动脉和颈动脉支架直径为5毫米,长13毫米)安装在高压球囊上,经皮置入10头猪的血管中[5个置于回旋支(CX),2个置于左前降支(LAD),5个置于肾动脉,3个置于颈动脉]。支架比天然血管直径大10 - 20%。所有动物在手术过程中接受5000国际单位肝素,并每日服用325毫克阿司匹林。在2、4、12、20、26和52周对动物进行随访血管造影和组织学检查。

结果

支架在荧光透视下易于显影,所有动物均成功置入,未发生球囊破裂或栓塞事件。无急性血栓形成事件。动物的随访血管造影显示所有肾动脉和颈动脉支架通畅,然而,动物中的2/7个冠状动脉支架在20周和52周时显示血管造影管腔狭窄(> 20%)。组织学检查显示支架部位有新生内膜形成,新生内膜平均厚度为325 - 650微米。

结论

在该动物模型中,这种支架是安全的,易于部署,具有出色的不透X线性,在不进行抗凝的情况下短期通畅性良好。正在进行临床试验和积累经验。