We have investigated the dependence of the wedge factors with field size, depth, nominal, and extended distances for 4, 6, 18, and 24 MV photon beams. Analysis of the experimental data suggests a general linear dependence of the wedge factors with field size and depth. The study shows that changes in wedge factors are insignificant (< or = +/-1.0%) with respect to measurements at nominal SSD, SAD, or extended SSD. This independence of the wedge factors on source-to-surface distance was studied for different photon energies (4-24 MV) and for different attenuating wedges (external and internal wedges). For clinical applications, an algorithm is presented to calculate the wedge factor dependence with field size and depth. The new algorithm has been successfully implemented to replace wedge look-up tables for dose and MU calculations in PRISM 1.2 treatment planning system used in our department.