Tailor R C, Followill D S, Hanson W F
Department of Radiation Physics, University of Texas M. D. Anderson Cancer Center, Houston 77030, USA.
Med Phys. 1998 Feb;25(2):241-4. doi: 10.1118/1.598187.
The Radiological Physics Center, through its dosimetry review visits to participating institutions, is aware that many institutions ignore the field-size and depth dependence of wedge transmission values. Reference wedge transmission values are normally measured by the Radiological Physics Center for a 10 cm x 10 cm field at the calibration depth of 5 or 7 cm. Recently, additional measurements (1) for a 10 cm x 10 cm field at 20-cm depth and (2) for a 20 cm x 20 cm field at the calibration depth were included. The transmission under these two conditions was compared with that under reference conditions. The relative transmission values for 138 photon beams from 88 separate linear accelerators (4-25 MV) and 60Co units were measured. Our data suggest that the dependence of the wedge transmission on field-size and depth, in the first approximation, depends on the absolute value of the transmission under reference conditions. For wedges with a transmission value greater than 0.65%, field-size dependence and change in depth dose are typically less than 2%. However, for wedges with transmission values less than 0.65%, field-size dependence increases with decreasing reference wedge transmission. The change in wedge transmission with depth is significant (> 2%) only for photon energies less than or equal to 10 MV and can exceed 5% for thick wedges. Failure to include the depth and field-size dependencies of wedge transmission in patient dosimetry calculations can result in significant tumor-dose discrepancies.
放射物理中心通过对参与机构的剂量学审查访问了解到,许多机构忽视了楔形滤过板透射值的射野大小和深度依赖性。参考楔形滤过板透射值通常由放射物理中心在5厘米或7厘米的校准深度下,针对10厘米×10厘米的射野进行测量。最近,增加了以下两项测量:(1)在20厘米深度下针对10厘米×10厘米的射野;(2)在校准深度下针对20厘米×20厘米的射野。将这两种情况下的透射与参考条件下的透射进行了比较。测量了来自88台独立直线加速器(4 - 25兆伏)和60钴装置的138束光子束的相对透射值。我们的数据表明,在一阶近似下,楔形滤过板透射对射野大小和深度的依赖性取决于参考条件下透射的绝对值。对于透射值大于0.65%的楔形滤过板,射野大小依赖性和深度剂量变化通常小于2%。然而,对于透射值小于0.65%的楔形滤过板,射野大小依赖性随参考楔形滤过板透射值的降低而增加。仅对于光子能量小于或等于10兆伏的情况,楔形滤过板透射随深度的变化才显著(>2%),对于厚楔形滤过板,该变化可能超过5%。在患者剂量学计算中未考虑楔形滤过板透射的深度和射野大小依赖性可能导致显著的肿瘤剂量差异。