Lee H, Buckley K M
University of Minnesota, Department of Electrical Engineering, Minneapolis 55455, USA.
IEEE Trans Biomed Eng. 1999 May;46(5):556-64. doi: 10.1109/10.759056.
A new electrocardiogram (ECG) data compression method is presented which employs a two dimensional (2-D) transform. This 2-D transform method utilizes the fact that ECG signals generally show two types of redundancies--between adjacent heartbeats and between adjacent samples. A heartbeat data sequence is cut and beat-aligned to form a 2-D data array. Any 2-D compression method can then be applied. Transform coding using the 2-D discrete cosine transform (DCT) [2-D DCT] is employed here as an example. Using selections from the MIT-BIH arrhythmia and Medtronic databases, results are presented that illustrate substantial improvement in compression ratio over one-dimensional methods for comparable percent root-mean-square difference (PRD).
提出了一种采用二维(2-D)变换的新型心电图(ECG)数据压缩方法。这种二维变换方法利用了心电图信号通常呈现两种冗余类型这一事实,即相邻心跳之间的冗余和相邻样本之间的冗余。将心跳数据序列进行切割并按心跳对齐,以形成二维数据阵列。然后可以应用任何二维压缩方法。这里以使用二维离散余弦变换(DCT)[二维DCT]的变换编码为例。使用从麻省理工学院 - 贝斯以色列女执事医疗中心(MIT - BIH)心律失常数据库和美敦力数据库中选取的数据,给出的结果表明,在可比的均方根差百分比(PRD)情况下,与一维方法相比,压缩比有显著提高。