Radial force of coronary stents: a comparative analysis.

作者信息

Rieu R, Barragan P, Masson C, Fuseri J, Garitey V, Silvestri M, Roquebert P, Sainsous J

机构信息

Institute of Mechanical Engineering, Technopole de Chateau Gambart, Marseilles, France.

出版信息

Catheter Cardiovasc Interv. 1999 Mar;46(3):380-91. doi: 10.1002/(SICI)1522-726X(199903)46:3<380::AID-CCD27>3.0.CO;2-J.

Abstract

High radial force has become an essential feature of new coronary stents. These stents are expected to exhibit sufficient radial force and minimum radial recoil when deployed in atheromatous lesions of various morphologies. The aim of this study was to compare the radial force of 17 coronary stents in vitro (Anglodynamics, ACS Multilink, AVE Micro II and GFX, Bard XT, Biocompatibles PC, Cook GR II, Cordis Crossflex, Hexacath Freedom, Johnson & Johnson PS 153'PS 154, and Crown, Medtronic Wiktor and BeStent BEL-15, Saint-Come SC 1616, Scimed Nir 7 and 9). Two in vitro mechanical tests were performed to assess the stent radial force. The first test measured the deformation of coronary stents in a V-stand by using a deformation controlled by a dynamometer based on a longitudinal generatrix. The precision was of 1 micron. This comparative test highlighted the appreciable variation in behavior of the range of tested stents in terms of resistance to local compression. The line graphs obtained show a very small (and in some cases nonexistent) purely elastic behavior area in the stents, with this result reflecting a small (<0.4 Newtons) range of forces applied to all stents. Above this value, the stents did not return to their size before compression and the line graph remains approximately linear for a long period. It was therefore possible to distinguish between two families of stents on the basis of the presence or absence of this elastic area. In the second test, the stent was deployed in a 3.0-mm elastic tube and a pressure gradient created between the interior and exterior of the tube. The precision was of 50 microns. A typical line graph of the pressure-diameter relationship was recognizable for a given stent. No difference in behavior between the different coronary stents was noted up to 0.3 x 10(5) Pa; subsequently, after 0.3 x 10(5) Pa, sizable deformation differences were visible between the least resistant stent, the BeStent BEL-15, and the most resistant stent, the Crossflex.

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验