Buitenweg J R, Rutten W L, Willems W P, van Nieuwkasteele J W
Institute of Biomedical Technology, Faculty of Electrical Engineering, University of Twente, Enschede, The Netherlands.
Med Biol Eng Comput. 1998 Sep;36(5):630-7. doi: 10.1007/BF02524436.
Sealing resistance is highly significant with respect to the electrical neuron-electrode contact because it decreases the stimulation threshold of neurons cultured on a planar micro-electrode array. A method is proposed for measurement of the sealing resistance using impedance spectroscopy. The effect of the sealing resistance on the total impedance spectrum of a cell-electrode interface is modelled for complete coverage of the electrode by the cell. Sensitivity analysis demonstrates that the impedance spectrum is determined by four parameters: two electrode parameters, the sealing resistance and the shunt capacitance between the lead of the electrode and the culture medium. Experimental verification of the model is performed by simultaneous measurement of the impedance spectrum and electrode coverage. A good and unique fit between the simulated and measured impedance spectra was obtained by varying the two electrode parameters and the sealing resistance.