Verellen D, Vinh-Hung V, Bijdekerke P, Nijs F, Linthout N, Bel A, Storme G
Department of Radiotherapy, Oncology Center, Academic Hospital-Free University, Brussels, (AZ-VUB), Belgium.
Radiother Oncol. 1999 Mar;50(3):355-66. doi: 10.1016/s0167-8140(99)00033-x.
The integration of a scanner for computed tomography (CT) and a treatment simulator (Sim-CT, Elekta Oncology Systems, Crawley, UK) has been studied in a clinical situation. Image quality, hounsfield units (HU) and linearity have been evaluated as well as the implications for treatment planning. The additional dose to the patient has also been highlighted.
Image data is acquired using an array of solid state X-ray detectors attached externally to the simulator's image intensifier. Three different fields of view (FOV: 25.0 cm, 35.0 cm and 50.0 cm) with 0.2 cm, 0.5 cm and 1.0 cm slice thickness can be selected and the system allows for an aperture diameter of 92.0 cm at standard isocentric height. The CT performance has been characterized with several criteria: spatial resolution, contrast sensitivity, geometric accuracy, reliability of hounsfield units and the radiation output level. The spatial resolution gauge of the nuclear associates quality phantom (NAQP) as well as modulation transfer functions (MTF) have been applied to evaluate the spatial resolution. Contrast sensitivity and HU measurements have been performed by means of the NAQP and a HU conversion phantom that allows inserts with different electron densities. The computed tomography dose index (CTDI) of the CT-option has been monitored with a pencil shaped ionization chamber. Treatment planning and dose calculations for heterogeneity correction based on the Sim-CT images generated from an anthropomorphic phantom as well as from ten patients have been compared with similar treatment plans based on identical, yet diagnostic CT (DCT) images.
The last row of holes that are resolved in the spatial resolution gauge of the NAQP are either 0.150 cm or 0.175 cm depending on the FOV and the applied reconstruction filter. These are consistent with the MTF curves showing cut-off frequencies ranging from 5.3 lp/cm to 7.1 lp/cm. Linear regression analysis of HU versus electron densities revealed a correlation coefficient of 0.99. Contrast, pixel size and geometric accuracy are within specifications. Computed tomography dose index values of 0.204 Gy/As and 0.069 Gy/As have been observed with dose measurements in the center of a 16 cm diameter and 32 cm diameter phantom, respectively for large FOV. Small FOV yields CTDI values of 0.925 Gy/As and 0.358 Gy/As which is a factor ten higher than the results obtained from a DCT under similar acquisition conditions. The phantom studies showed excellent agreement between dose distributions generated with the Sim-CT and DCT HU. The deviations between the calculated settings of monitor units as well as the maximum dose in three dimensions were less than 1% for the treatment plans based on either of these HU both for pelvic as well as thoracic simulations. The patient studies confirmed these results.
The CT-option can be considered as an added value to the simulation process and the images acquired on the Sim-CT system are adequate for dose calculation with tissue heterogeneity correction. The good image quality, however, is compromised by the relative high dose values to the patient. The considerable load to the conventional X-ray tube currently limits the Sim-CT to seven image acquisitions per patient and therefore the system is limited in its capability to perform full three-dimensional reconstruction.
已在临床环境中研究了计算机断层扫描(CT)扫描仪与治疗模拟器(Sim-CT,医科达肿瘤系统公司,英国克劳利)的整合情况。已对图像质量、亨氏单位(HU)和线性度进行了评估,以及对治疗计划的影响。还强调了对患者的额外剂量。
使用外部连接到模拟器图像增强器的固态X射线探测器阵列获取图像数据。可以选择三种不同的视野(FOV:25.0厘米、35.0厘米和50.0厘米),切片厚度分别为0.2厘米、0.5厘米和1.0厘米,并且该系统在标准等中心高度允许孔径直径为92.0厘米。CT性能已通过多个标准进行表征:空间分辨率、对比度灵敏度、几何精度、亨氏单位的可靠性和辐射输出水平。已应用核关联质量体模(NAQP)的空间分辨率测量仪以及调制传递函数(MTF)来评估空间分辨率。对比度灵敏度和HU测量已通过NAQP以及一个允许插入具有不同电子密度物体的HU转换体模进行。已使用笔形电离室监测CT选项的计算机断层扫描剂量指数(CTDI)。已将基于从拟人化体模以及十名患者生成的Sim-CT图像进行的用于不均匀性校正的治疗计划和剂量计算,与基于相同但为诊断性CT(DCT)图像的类似治疗计划进行了比较。
NAQP空间分辨率测量仪中可分辨的最后一排孔,根据视野和应用的重建滤波器,要么是0.150厘米,要么是0.175厘米。这些与MTF曲线一致,MTF曲线显示截止频率范围为5.3线对/厘米至7.1线对/厘米。HU与电子密度的线性回归分析显示相关系数为0.99。对比度、像素大小和几何精度均在规格范围内。对于大视野,在直径16厘米和32厘米体模的中心进行剂量测量时,分别观察到CTDI值为0.204戈瑞/安秒和0.069戈瑞/安秒。小视野产生的CTDI值为0.925戈瑞/安秒和0.358戈瑞/安秒,这比在类似采集条件下从DCT获得的结果高十倍。体模研究表明,用Sim-CT和DCT HU生成的剂量分布之间具有极好的一致性。基于这两种HU中的任何一种进行的盆腔以及胸部模拟的治疗计划,在监测单位的计算设置以及三维最大剂量之间的偏差均小于1%。患者研究证实了这些结果。
CT选项可被视为模拟过程的一项附加价值,并且在Sim-CT系统上获取的图像足以用于进行组织不均匀性校正的剂量计算。然而,良好的图像质量因对患者的相对高剂量值而受到影响。当前传统X射线管的相当大负荷将Sim-CT限制为每位患者只能进行七次图像采集,因此该系统在进行完整三维重建的能力方面受到限制。