Tono-Pen tonometer and corneal thickness.

作者信息

Mok K H, Wong C S, Lee V W

机构信息

Queen Mary Hospital Unit, Department of Ophthalmology and Visual Sciences, Chinese University of Hong Kong, China.

出版信息

Eye (Lond). 1999;13 ( Pt 1):35-7. doi: 10.1038/eye.1999.7.

Abstract

PURPOSE

To compare the intraocular pressure (IOP) measurements taken at the central cornea with those taken at the mid-peripheral cornea using the Tono-Pen tonometer.

METHODS

Forty-eight normal human subjects had their IOP measurements taken (i) with the Goldmann tonometer, (ii) with the Tono-Pen at the central cornea and (iii) with the Tono-Pen at the mid-peripheral cornea. The central and mid-peripheral corneal thickness at the site of IOP measurements were then determined by ultrasonic pachymetry.

RESULTS

The mean IOP with the Goldmann tonometer was 15.7 +/- 3.1 mmHg. The mean Tono-Pen IOPs at the central and midperipheral cornea were 15.2 +/- 3.2 mmHg and 15.7 +/- 3.3 mmHg respectively. The mean midperipheral corneal thickness (578 microns) was significantly higher than the mean central thickness (538 microns) (p < 0.001).

CONCLUSION

There was no significant difference between the IOP taken with the Tono-Pen and with the Goldmann tonometer. In addition, no clinically significant difference was observed between the IOP readings of central and mid-peripheral cornea measured by the Tono-Pen.

摘要

目的

使用眼压计比较在中央角膜和周边角膜中部测量的眼压(IOP)。

方法

48名正常受试者分别使用(i)Goldmann眼压计、(ii)眼压计在中央角膜测量眼压、(iii)眼压计在周边角膜中部测量眼压。然后通过超声测厚法测定眼压测量部位的中央和周边角膜厚度。

结果

Goldmann眼压计测量的平均眼压为15.7±3.1 mmHg。眼压计在中央角膜和周边角膜中部测量的平均眼压分别为15.2±3.2 mmHg和15.7±3.3 mmHg。周边角膜中部的平均厚度(578微米)显著高于中央平均厚度(538微米)(p<0.001)。

结论

眼压计测量的眼压与Goldmann眼压计测量的眼压之间无显著差异。此外,眼压计测量的中央角膜和周边角膜中部的眼压读数之间未观察到临床显著差异。