Curr Eye Res. 2021 Jan;46(1):71-77. doi: 10.1080/02713683.2020.1782942. Epub 2020 Jul 3.
To assess the validity of and compare applanation and rebound tonometry readings of intraocular pressure in alert normal chicks from ages 3 to 45 days.
Intraocular pressures (IOPs) were measured weekly in awake White Leghorn chicks, from ages 3-45 days (n = 22-30 per age group), with both applanation Tono-Pen and rebound TonoLab tonometers. Three repeated measurements on individual eyes were used to derive variance data for both instruments at each age. Calibration curves were also derived for each instrument and each age, weekly from ages 10-45 days (n = 3-4 per age group), from manometry data collected over IOP settings of 0 to 100 mmHg in 5 mmHg steps in cannulated eyes.
The TonoLab showed less within measurement variability, but more variability with age, than the Tono-Pen. The coefficient of variation ranged from 3.8-8.3% for the TonoLab, compared to 11.0-19.7% for the Tono-Pen across all ages. For the youngest, 3 day-old chicks, mean IOPs recorded with the Tono-Pen and TonoLab were not significantly different (17.0 ± 5.6 and 15.2 ± 3.7 mmHg, respectively, = .27). However, with increasing age, IOP readings significantly increased for the TonoLab ( < .001), whereas Tono-Pen readings did not. Compared to manometry settings, the Tono-Pen tended to underestimate IOPs while the TonoLab overestimated IOPs over the range 20-60 mmHg, saturating thereafter; there were also age-dependent differences for the TonoLab.
Both the Tono-Pen and TonoLab gave IOP readings that differed from manometry settings in normal young chicks over some or all of the ages tested. These results reinforce the importance of calibrating clinical tonometers in animal studies involving IOP as a key variable.
评估压平眼压计和回弹眼压计在 3 至 45 日龄警觉正常鸡眼中测量眼压的有效性并进行比较。
每周在 3 至 45 日龄(每组 22-30 只)的白来航鸡清醒状态下用压平眼压计 Tono-Pen 和回弹眼压计 TonoLab 测量眼内压。使用两种仪器在每个年龄段的三只重复测量眼上获得方差数据。在插管眼内,从 10 至 45 日龄(每组 3-4 只),每周根据眼压设定 0 至 100mmHg 每隔 5mmHg 一次的眼压计数据,为两种仪器和每个年龄段都推导出校准曲线。
TonoLab 的测量内变异性较小,但随年龄变化的变异性大于 Tono-Pen。TonoLab 的变异系数范围为 3.8-8.3%,而 Tono-Pen 则为 11.0-19.7%,所有年龄段均如此。对于最年轻的 3 日龄小鸡,Tono-Pen 和 TonoLab 记录的平均眼压没有显著差异(分别为 17.0 ± 5.6mmHg 和 15.2 ± 3.7mmHg, =.27)。然而,随着年龄的增长,TonoLab 的眼压读数显著增加(<.001),而 Tono-Pen 则没有。与眼压计设定值相比,Tono-Pen 倾向于低估眼压,而 TonoLab 则在 20-60mmHg 范围内高估眼压,此后趋于饱和;TonoLab 也存在与年龄相关的差异。
在一些或所有测试年龄段的正常雏鸡中,Tono-Pen 和 TonoLab 的眼压读数均与眼压计设定值不同。这些结果强调了在涉及眼压作为关键变量的动物研究中校准临床眼压计的重要性。