Bogart D B, Jung S C
Northland Cardiology, North Kansas City, Missouri 64116, USA.
Catheter Cardiovasc Interv. 1999 Jul;47(3):323-4. doi: 10.1002/(sici)1522-726x(199907)47:3<323::aid-ccd16>3.0.co;2-p.
We report a technique for retrieval of a dislodged coronary stent using a stiff angioplasty wire positioned beside the initial stent guidewire. This two-wire technique provides a better platform to move and position the snare device without moving the dislodged stent and thus lessens the risk of embolization. If a larger femoral sheath is needed for stent removal, this method facilitates sheath exchange. Cathet. Cardiovasc. Intervent. 47:323-324, 1999.