Li Chunjian, Yang Zhijian, Cao Kejiang
Department of Cardiology, the First Affiliated Hospital of Nanjing Medical University, Nanjing, Jiangsu 210029, China.
J Biomed Res. 2010 Nov;24(6):479-82. doi: 10.1016/S1674-8301(10)60064-4.
A rapamycin-eluting stent was dislodged during attempt of implantation at the proximal right coronary artery, which was found by fluoroscopy to have migrated into the anterior trunk of the left renal artery. We chose a 5 mm diameter Amplatz gooseneck snare and successfully retrieved the lost stent from the lodging vessel.
一枚雷帕霉素洗脱支架在右冠状动脉近端植入尝试过程中发生移位,经荧光透视发现其已迁移至左肾动脉前干。我们选用了直径5毫米的Amplatz鹅颈圈套器,并成功从栓塞血管中取出了丢失的支架。