Suppr超能文献

颞下颌关节假体的磨损测试:超高分子量聚乙烯(UHMWPE)和聚四氟乙烯(PTFE)与金属球在水和血清中的磨损测试

Wear-testing of a temporomandibular joint prosthesis: UHMWPE and PTFE against a metal ball, in water and in serum.

作者信息

Van Loon J P, Verkerke G J, de Bont L G, Liem R S

机构信息

Department of Oral & Maxillofacial Surgery, University Hospital Groningen, The Netherlands.

出版信息

Biomaterials. 1999 Aug;20(16):1471-8. doi: 10.1016/s0142-9612(99)00042-3.

Abstract

For a temporomandibular joint prosthesis, an estimation of the wear rate was needed, prior to patient application. Therefore, we determined the in vitro wear rate of the ball-socket articulation of this prosthesis, consisting of a metal head and an ultra-high molecular weight polyethylene (UHMWPE) cup. The basic testing configuration consisted of one 8-mm diameter stainless-steel ball, rotating between two conforming cups with a minimum thickness of 5 mm. For validation of the testing apparatus, two cup materials, in two lubricants, were tested. Both cup materials, UHMWPE and polytetrafluoroethylene (PTFE) were tested in deionized water, as well as in a serum-based solution. For UHMWPE in serum, eight samples were tested, for the other combinations four samples. For UHMWPE, the tests ran for 7 million cycles, for PTFE between 0.8 and 1.7 million cycles. For UHMWPE, the wear rate was 0.006 and 0.47 (mm3/10(6) cycles), in water and in serum, respectively. For PTFE, the wear rate was 2.8 and 47 (mm3/10(6) cycles), in water and in serum, respectively. For reason that testing in serum simulates the in vivo situation best, it was concluded that the wear rate of the TMJ prosthesis articulation is 0.47 (mm3/10(6) cycles), which is considered acceptable.

摘要

对于颞下颌关节假体,在应用于患者之前需要估计其磨损率。因此,我们测定了该假体球窝关节的体外磨损率,该关节由金属头和超高分子量聚乙烯(UHMWPE)杯组成。基本测试配置包括一个直径8毫米的不锈钢球,在两个最小厚度为5毫米的适配杯中旋转。为了验证测试设备,对两种杯材料在两种润滑剂中进行了测试。两种杯材料,即UHMWPE和聚四氟乙烯(PTFE),分别在去离子水中和基于血清的溶液中进行了测试。对于血清中的UHMWPE,测试了8个样本,对于其他组合测试了4个样本。对于UHMWPE,测试进行了700万次循环,对于PTFE进行了0.8至170万次循环。对于UHMWPE,在水中和血清中的磨损率分别为0.006和0.47(mm³/10⁶循环)。对于PTFE,在水中和血清中的磨损率分别为2.8和47(mm³/10⁶循环)。由于在血清中的测试最能模拟体内情况,得出结论:颞下颌关节假体关节的磨损率为0.47(mm³/10⁶循环),这被认为是可以接受的。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验