新型跟骨定量超声系统的评估及中国南方女性超声正常值的测定。

Evaluation of a new calcaneal quantitative ultrasound system and determination of normative ultrasound values in southern Chinese women.

作者信息

Kung A W, Tang G W, Luk K D, Chu L W

机构信息

Department of Medicine, University of Hong Kong, Queen Mary Hospital, Hong Kong.

出版信息

Osteoporos Int. 1999;9(4):312-7. doi: 10.1007/s001980050153.

Abstract

Quantitative ultrasound (QUS) assessment at the calcaneus has been found to be a safe and reliable method for evaluating skeletal status. The present study aimed at evaluating the precision of the Sahara bone ultrasound densitometer and to determine the normative QUS data in healthy southern Chinese women. Broadband ultrasound attenuation (BUA), speed of sound (SOS) and qualitative ultrasound index (QUI) were determined. The long-term in vitro precision of the Sahara machine over 6 months was 4.6% for BUA and 0.39% for SOS. The short-term in vivo precision was 3.2 +/- 1.3% for BUA, 0.3 +/- 0.2% for SOS and 1.8 +/- 1.0% for QUI. The standardized precision for BUA, SOS and QUI was 4.4, 3.8 and 2.2 respectively. The normative data were determined in 1086 healthy subjects. Postmenopausal women had significantly lower BUA, SOS and QUI levels than the premenopausal women. Significant negative correlations were observed between QUS indices and age. Bone mineral density (BMD) assessments was performed on 349 of these subjects. BUA correlated significantly with lumbar spine BMD (r = 0.326) and femoral neck BMD (r = 0.395). Similar correlations were observed between SOS, QUI and BMD, with r values ranging between 0.446 to 0.522. Despite the fact that Chinese women have significantly lower BMD values than Caucasian women, the mean BUA values for pre- and postmenopausal Chinese women (73 +/- 18 and 59 +/- 18 dB/MHz respectively) were almost the same as those reported for Caucasian women. These normative data will be useful in the assessment of southern Chinese women with fracture risk.

摘要

跟骨定量超声(QUS)评估已被发现是一种评估骨骼状况的安全可靠方法。本研究旨在评估Sahara骨超声密度仪的精度,并确定中国南方健康女性的QUS规范数据。测定了宽带超声衰减(BUA)、声速(SOS)和定性超声指数(QUI)。Sahara仪器在6个月内的长期体外精度,BUA为4.6%,SOS为0.39%。短期体内精度,BUA为3.2±1.3%,SOS为0.3±0.2%,QUI为1.8±1.0%。BUA、SOS和QUI的标准化精度分别为4.4、3.8和2.2。在1086名健康受试者中确定了规范数据。绝经后女性的BUA、SOS和QUI水平显著低于绝经前女性。QUS指标与年龄之间存在显著负相关。对其中349名受试者进行了骨密度(BMD)评估。BUA与腰椎BMD(r = 0.326)和股骨颈BMD(r = 0.395)显著相关。在SOS、QUI和BMD之间也观察到类似的相关性,r值在0.446至0.522之间。尽管中国女性的BMD值显著低于白种女性,但绝经前和绝经后中国女性的平均BUA值(分别为73±18和59±18 dB/MHz)与白种女性报告的数值几乎相同。这些规范数据将有助于评估有骨折风险的中国南方女性。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验