An in vitro comparison of bone deformation measured with surface and staple mounted strain gauges.

作者信息

Arndt A, Westblad P, Ekenman I, Halvorsen K, Lundberg A

机构信息

Department of Orthopaedic Surgery, Huddinge University Hospital, Sweden.

出版信息

J Biomech. 1999 Dec;32(12):1359-63. doi: 10.1016/s0021-9290(99)00129-3.

Abstract

Chicken tibiae were chosen as a model for human second metatarsals. Local surface bone deformation in a 4-point bending configuration was measured in vitro by both strain gauge instrumented staples and strain gauges bonded to the bone's cortical surface. A series of staple bridge dimensions (0.5, 0.6, 0.8 and 1.0 mm) was compared to test for staple influence on bone characteristics and greatest measurement validity and reliability. Thicker staple inhibition of bone deformation was the greatest but differences to thinner staples were not statistically significant (p > 0.05). All staples except 0.5 mm had maximum deviations from linearity less than 1%. The 1.0 mm staple had an R2 value of 0.992 +/- 0.006 plotted against the 4-point bending input force and 0.994 +/- 0.002 plotted against the surface strain gauge signal. The mean intraclass correlation coefficients (ICC) calculated with four input forces (30, 60, 90 and 120 N) and for loading and unloading conditions for the 0.5, 0.6, 0.8 and 1.0 mm staples were 0.75, 0.83, 0.87 and 0.92, respectively. Finally, the differences in slope of the staple strain gauge signal plotted against surface strain gauge signal between input force loading and unloading conditions (0.32), and between input compression and tension conditions (0.79) was least for the 1.0 mm staple which also resulted in the lowest standard deviations. These results suggested the appropriateness of the 1.0 mm staple for in vivo application.

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验