Wallgren C G, Hanson J S, Kretzschmar G, Zetterqvist P
Acta Paediatr Scand Suppl. 1975(254):1-6. doi: 10.1111/j.1651-2227.1975.tb15005.x.
Cardiac output was assessed in 16 children without evidence of cardiovascular shunts from simultaneously recorded earpiece and cuvette densitometric tracings. The earpiece set-up was calibrated by the end-tail method. Both instruments gave linear response to dye concentrations used. There was a close correlation between cardiac output figures recorded by the two methods (r equal to 0.95) and no systematic difference occurred. It is concluded that in infants and children earpiece densitometry constitutes a rapid and accurate method for quantitative circulatory studies.
通过同时记录耳塞和比色计密度描记图,对16名无心血管分流证据的儿童的心输出量进行了评估。耳塞装置采用尾端法进行校准。两种仪器对所用染料浓度均呈线性反应。两种方法记录的心输出量数据之间存在密切相关性(r等于0.95),且未出现系统差异。结论是,在婴幼儿中,耳塞密度测定法是一种用于定量循环研究的快速且准确的方法。