Measurement of cardiac output by earpiece dye-dilution method with automatic calibration of dye concentration.

作者信息

Katori R, Hayashi T, Kanamasa K, Ishikawa K

出版信息

Tohoku J Exp Med. 1977 May;122(1):51-8. doi: 10.1620/tjem.122.51.

Abstract

A non-invasive method for measuring cardiac output by an earpiece dye densitometer was proposed. The densitometer is dichromatic and has an air capsule to make subject's ear bloodless by inflation, so that it can calibrate indocyanine green dye concentration without blood sampling. Duplicate measurements of cardiac output showed a good agreement in 40 cases (r = 0.97, standard deviation (S.D.) = 8.9%), which was comparable to the result of the cuvette method (r = 0.98, S.D. = 8.1%). Simultaneous measurements of cardiac output by the earpiece method (x) and the standard cuvette method (y) revealed a good agreement (r = 0.91, y = 0.96x+0.34, S.D. = 16.4%) in 52 measurements of 25 cases. A similarly good correlation was obtained between the two methods during ergometer exercise in supine position in 5 cases. These suggest that this earpiece dye-dilution method is reliable for cardiac output measurement and advantageous for clinical use because of non-invasive technique.

摘要

提出了一种通过耳塞式染料密度计测量心输出量的非侵入性方法。该密度计是双色的,并有一个气囊,通过充气使受试者耳部无血,从而无需采血即可校准吲哚菁绿染料浓度。对40例患者的心输出量进行重复测量,结果显示一致性良好(r = 0.97,标准差(S.D.)= 8.9%),这与比色皿法的结果(r = 0.98,S.D. = 8.1%)相当。对25例患者进行52次测量,通过耳塞法(x)和标准比色皿法(y)同时测量心输出量,结果显示一致性良好(r = 0.91,y = 0.96x + 0.34,S.D. = 16.4%)。对5例患者在仰卧位进行测力计运动期间,两种方法之间也获得了类似的良好相关性。这些结果表明,这种耳塞式染料稀释法在心输出量测量方面是可靠的,并且由于其非侵入性技术,在临床应用中具有优势。

相似文献

4
Fiberoptic ear densitometer for measurement of cardiac output.
Can J Physiol Pharmacol. 1978 Feb;56(1):69-75. doi: 10.1139/y78-010.
5
A calibration device for the earpiece dichromatic densitometer.
J Appl Physiol. 1968 Nov;25(5):633-5. doi: 10.1152/jappl.1968.25.5.633.