Beekman W H, Hage J J, Taets van Amerongen A H, Mulder J W
Department of Plastic and Reconstructive Surgery, Academisch Ziekenhuis Vrije Universiteit, Amsterdam, The Netherlands.
Scand J Plast Reconstr Surg Hand Surg. 1999 Dec;33(4):415-8. doi: 10.1080/02844319950159127.
We prospectively studied the accuracy of magnetic resonance imaging (MRI) and ultrasonography (US) for preoperative detection of rupture in 35 single-lumen implants filled with silicone gel in 18 patients. The positive predictive value of US for rupture of an implant was 70% and the negative predictive value 64%. Sensitivity and specificity were 44% and 87%, respectively. Accuracy, defined as the total true positive and true negative values divided by the total number of implants studied was 66%. The positive predictive value of MRI was 100% and the negative predictive value 90%. The corresponding sensitivity and specificity were 88% and 100% and the accuracy 94%. MRI offers significantly better diagnostic sensitivity (p = 0.02) and accuracy (p = 0.004), and should be regarded as the "gold standard" in the evaluation of rupture of breast implants filled with silicone gel. When MRI is not readily available, US is an acceptable alternative.
我们前瞻性地研究了磁共振成像(MRI)和超声检查(US)对18例患者体内35个单腔硅胶填充植入物术前破裂检测的准确性。超声检查对植入物破裂的阳性预测值为70%,阴性预测值为64%。敏感性和特异性分别为44%和87%。准确性定义为真阳性和真阴性值总数除以所研究植入物的总数,为66%。MRI的阳性预测值为100%,阴性预测值为90%。相应的敏感性和特异性分别为88%和100%,准确性为94%。MRI具有显著更好的诊断敏感性(p = 0.02)和准确性(p = 0.004),应被视为评估硅胶填充乳房植入物破裂的“金标准”。当无法轻易获得MRI时,超声检查是一种可接受的替代方法。