骨支撑对骨折萎缩无牙下颌骨模型固定强度影响的体外研究。

An in vitro study of the effect of bony buttressing on fixation strength of a fractured atrophic edentulous mandible model.

作者信息

Sikes J W, Smith B R, Mukherjee D P

机构信息

Division of Oral and Maxillofacial Surgery, Louisiana State University Medical Center at Shreveport, LA 71130, USA.

出版信息

J Oral Maxillofac Surg. 2000 Jan;58(1):56-61; discussion 62. doi: 10.1016/s0278-2391(00)80017-8.

Abstract

PURPOSE

The purpose of this study was to measure the resistance to displacement in an adult bovine rib mandible model as the vertical dimension of the bone decreases, and to compare the fixation strength of titanium miniplates and reconstruction plates.

MATERIALS AND METHODS

Five groups of 6 ribs each were tested based on the vertical dimension of the rib and method of fixation (group 1, 40 mm, miniplate), (group 2, 30 mm, miniplate), (group 3, 20 mm, miniplate), (group 4, 10 mm, miniplate), (group 5, 10 mm, reconstruction plate). In the 4 groups stabilized with a miniplate, a 3-hole-per-segment, 2.0-mm titanium miniplate and 6.0-mm monocortical titanium screws were used. In group 5, a 3-screw-per-segment, 2.4-mm titanium reconstruction plate and 2.4-mm bicortical titanium screws were used for fixation. A 2-dimensional beam model (Class III lever) was established, and an Instron machine was used initially to develop a load-displacement curve to 100 N for each specimen. An osteotomy was then created, and the segments were reduced without preload. The ribs were then loaded to failure in the Instron machine. The load-displacement curve and failure load were recorded.

RESULTS

There was no significant difference (P < or = .05) between any of the groups before the osteotomy. At 75 N, groups 1 and 2 were superior to group 5 in resistance to displacement. Group 4 showed significantly less (P < or = .05) resistance to displacement than the other experimental miniplated subgroups. As loads increased, groups 1 and 2 continued to show increased resistance to displacement when compared with groups 3, 4, and 5. Loads to failure for groups 1 and 2 were comparable to group 5.

CONCLUSIONS

Decreased resistance to displacement occurs with decreasing vertical dimension of the bovine rib. In fractures of the edentulous mandible, a miniplate is more likely to provide adequate fixation if the mandible is 30 to 40 mm in height (nonatrophic). At higher loads, groups with greater vertical height (30 mm and 40 mm) provided resistance to displacement equivalent to the 10-mm group repaired with a reconstruction plate. Therefore, fractures of the atrophic mandible may be better treated with more rigid techniques.

摘要

目的

本研究的目的是在成年牛肋骨下颌骨模型中,测量随着骨垂直尺寸减小的移位阻力,并比较微型钛板和重建板的固定强度。

材料与方法

根据肋骨的垂直尺寸和固定方法将五组,每组6根肋骨进行测试(第1组,40mm,微型钛板),(第2组,30mm,微型钛板),(第3组,20mm,微型钛板),(第4组,10mm,微型钛板),(第5组,10mm,重建板)。在使用微型钛板固定的4组中,使用每段3孔、2.0mm的微型钛板和6.0mm的单皮质钛螺钉。在第5组中,使用每段3枚螺钉、2.4mm的钛重建板和2.4mm的双皮质钛螺钉进行固定。建立二维梁模型(III类杠杆),最初使用英斯特朗材料试验机为每个标本绘制至100N的载荷-位移曲线。然后进行截骨术,在无预载的情况下复位骨段。然后在英斯特朗材料试验机中对肋骨加载直至破坏。记录载荷-位移曲线和破坏载荷。

结果

截骨术前任何组之间均无显著差异(P≤0.05)。在75N时,第1组和第2组在移位阻力方面优于第5组。第4组的移位阻力明显小于其他微型钛板固定的实验亚组(P≤0.05)。随着载荷增加,与第3、4和5组相比,第1组和第2组继续表现出增加的移位阻力。第1组和第2组的破坏载荷与第5组相当。

结论

随着牛肋骨垂直尺寸减小,移位阻力降低。在无牙下颌骨骨折中,如果下颌骨高度为30至40mm(非萎缩性),微型钛板更有可能提供足够的固定。在较高载荷下,垂直高度较大(30mm和40mm)的组提供的移位阻力与用重建板修复的10mm组相当。因此,萎缩性下颌骨骨折可能用更坚固的技术治疗效果更好。