Assessment of blood lactate: practical evaluation of the Biosen 5030 lactate analyzer.

作者信息

Davison R C, Coleman D, Balmer J, Nunn M, Theakston S, Burrows M, Bird S

机构信息

Department of Sport Science, Canterbury Christ Church University College, United Kingdom.

出版信息

Med Sci Sports Exerc. 2000 Jan;32(1):243-7. doi: 10.1097/00005768-200001000-00036.

Abstract

PURPOSE

The aim of this study was to assess the validity and reliability of the Biosen 5030 lactate analyzer compared with a YSI 2300 lactate analyzer and a Kodak Ektachem DTII in a practical laboratory study context.

METHODS

To assess validity, 144 triplicate capillarized blood samples, across a range of values, were analyzed using the three analyzers. To assess reliability a further 665 samples were repeat analyzed. Temporal stability was determined by the reanalysis of resting and maximal exercise blood samples, after a period of storage ranging from 7 to 20 h, at room temperature. To measure inter- and intra-investigator reliability, 20 resting samples were taken from three different subjects by different investigators and a coefficient of variation was determined.

RESULTS

There were strong relationships between the Biosen, the YSI (r2 = 0.97), and the Kodak Ektachem (r2 = 0.91). An analysis of Biosen compared with YSI revealed a positive bias of 0.37 mmol x L(-1) (95% limits of agreement, -0.85 to 1.59 mmol x L(-1)). The test-retest reliability correlation was significant (r2 = 0.99, P < 0.05), but a paired t-test revealed a small (0.03 mmol x L(-1), P < 0.05) significant difference. The coefficient of variation from the three investigators across the 20 samples ranged from 1.3 to 3%. Blood lactate concentration in resting blood samples did significantly increase in value (0.2 mmol x L(-1), P < 0.05) after 7-h exposure to the air, whereas there was no change in maximal exercise blood lactate values after 20-h exposure to the air.

CONCLUSIONS

In a practical context, the Biosen 5030 lactate analyzer was comparable to the other analyzers giving fast reliable measures of blood lactate concentrations over the full range of values, which remained stable over extended periods at room temperature.

摘要

目的

本研究旨在在实际实验室研究环境中,评估Biosen 5030乳酸分析仪与YSI 2300乳酸分析仪和柯达Ektachem DTII相比的有效性和可靠性。

方法

为评估有效性,使用这三种分析仪对144份不同值的毛细血管血样本进行了三次重复分析。为评估可靠性,又对665份样本进行了重复分析。通过在室温下储存7至20小时后,对静息和最大运动血样本进行重新分析来确定时间稳定性。为测量研究者间和研究者内的可靠性,不同研究者从三名不同受试者身上采集了20份静息样本,并确定了变异系数。

结果

Biosen与YSI(r2 = 0.97)以及柯达Ektachem(r2 = 0.91)之间存在很强的相关性。将Biosen与YSI进行比较分析发现,存在0.37 mmol x L(-1)的正偏差(95%一致性界限,-0.85至1.59 mmol x L(-1))。重测可靠性相关性显著(r2 = 0.99,P < 0.05),但配对t检验显示存在微小的(0.03 mmol x L(-1),P < 0.05)显著差异。三位研究者对20份样本的变异系数范围为1.3%至3%。静息血样本在暴露于空气中7小时后,血乳酸浓度确实显著升高(0.2 mmol x L(-1),P < 0.05),而最大运动血乳酸值在暴露于空气中20小时后没有变化。

结论

在实际环境中,Biosen 5030乳酸分析仪与其他分析仪相当,能够在整个值范围内快速可靠地测量血乳酸浓度,并且在室温下长时间保持稳定。