Eur J Appl Physiol. 2024 Dec;124(12):3781-3788. doi: 10.1007/s00421-024-05572-6. Epub 2024 Aug 19.
To evaluate the precision and accuracy in measured blood lactate concentrations among four commonly used handheld lactate analyzers compared to two stationary analyzers.
Venous blood samples were taken at exercise intensities ranging from low to high. The blood lactate concentration was measured simultaneously with four pairs of handheld lactate analyzers (two new units of each brand: Lactate Plus, Lactate Pro2, Lactate Scout 4, and TaiDoc TD-4289), and compared with two stationary analyzers (Biosen C-Line and YSI Sport 1500). Measurements were repeated for a range of blood lactate concentrations (measured with Biosen) from 0.88 to 4.89 mM with a median difference between measurements of 0.10 mM.
The mean relative differences to the Biosen analyzer were 7% (Plus), 7% (Pro), 10% (Scout), 42% (Tai), and 32% (Ysi). The residual standard errors after linear regression against Biosen were 0.18 mM (Plus), 0.20 mM (Pro), 0.22 mM (Scout), 0.15 mM (Tai), and 0.06 mM (Ysi). Accordingly, a blood lactate concentration of 3 mM measured with Biosen yielded 95% prediction intervals that were 0.72 mM (Plus), 0.80 mM (Pro), 0.87 mM (Scout), 0.60 mM (Tai), and 0.23 mM (Ysi) wide.
Compared to our two stationary analyzers, the precision of the four handheld lactate analyzers evaluated in this study was poor. Among the four, Tai was the most precise; however, this analyzer had low accuracy with a substantial mean difference to the reference analyzer.
评估四种常用手持式乳酸分析仪与两种固定式分析仪相比,在测量血液乳酸浓度方面的精度和准确性。
在低至高强度的运动强度下采集静脉血样本。同时使用四对手持式乳酸分析仪(每个品牌的两个新单位:Lactate Plus、Lactate Pro2、Lactate Scout 4 和 TaiDoc TD-4289)以及两种固定式分析仪(Biosen C-Line 和 YSI Sport 1500)测量血液乳酸浓度。测量范围为 0.88 至 4.89 mM,两次测量的中位数差值为 0.10 mM。
与 Biosen 分析仪相比,平均相对差异分别为 7%(Plus)、7%(Pro)、10%(Scout)、42%(Tai)和 32%(Ysi)。线性回归后与 Biosen 的残差标准误差分别为 0.18 mM(Plus)、0.20 mM(Pro)、0.22 mM(Scout)、0.15 mM(Tai)和 0.06 mM(Ysi)。因此,用 Biosen 测量的 3 mM 血液乳酸浓度产生的 95%预测区间分别为 0.72 mM(Plus)、0.80 mM(Pro)、0.87 mM(Scout)、0.60 mM(Tai)和 0.23 mM(Ysi)。
与我们的两种固定式分析仪相比,本研究评估的四种手持式乳酸分析仪的精度较差。在这四种分析仪中,Tai 最精确;然而,该分析仪的准确性较低,与参考分析仪存在较大的平均差异。