将Gianturco弹簧圈栓塞入肺动脉,需要紧急外科干预。

Embolization of Gianturco coil into the pulmonary artery requiring emergency surgical intervention.

作者信息

Hijazi A, Mazhar R, Bricelj V, Robida A

机构信息

Consultant Cardiovascular-Thoracic Surgery Department, Hamad Medical Corporation, Doha, Qatar.

出版信息

Tex Heart Inst J. 1999;26(4):300-2.

Abstract

We report the case of a 20-month-old girl who underwent Gianturco coil embolization to a patent ductus arteriosus in May 1997. The coil migrated to the pulmonary artery. After unsuccessful attempts to retrieve it with snares and forceps, we engaged the coil with an end-hole balloon catheter and pulled it down to the right ventricle. There it became entangled in the tricuspid valvular apparatus and could not be moved farther. Due to concerns about sequelae, the patient was referred for surgery. Following a mid-sternotomy under cardiopulmonary bypass, we removed the coil and ligated the patent ductus arteriosus. The patient made an uneventful recovery. A brief review of the literature is presented.

摘要

我们报告了一例20个月大女孩的病例,该女孩于1997年5月接受了动脉导管未闭的Gianturco弹簧圈栓塞术。弹簧圈迁移至肺动脉。在用圈套器和镊子尝试取出失败后,我们用端孔球囊导管套住弹簧圈并将其拉至右心室。在那里,它缠绕在三尖瓣装置中,无法再移动。由于担心后遗症,该患者被转诊接受手术。在体外循环下进行胸骨正中切开术后,我们取出了弹簧圈并结扎了动脉导管未闭。患者恢复顺利。本文还对相关文献进行了简要综述。