Berndt A, Rickey D W, Rathee S, Bews J
Cancercare Manitoba, Winnipeg, Canada.
Med Phys. 2000 Jan;27(1):129-31. doi: 10.1118/1.598874.
In this paper we describe techniques for measuring the dimensions and position reproducibility of an 192Ir brachytherapy stepping source. Measurements were carried out using a 0.25x10x152 mm3 collimator placed in front of a detector of our own design. The brachytherapy source was translated past the collimator in 0.025 mm increments using a stepper motor. The source was found to be 3.58 mm long and 0.69 mm wide, which is in good agreement with the manufacturer's values of 3.5x0.6 mm2. The source position was reproducible to within 0.12 mm.