Absorbed dose determination at short distance from 60Co and 192Ir brachytherapy sources.

作者信息

Tölli H, Johansson K A

机构信息

Department of Radiation Physics, University of Göteborg, Sahlgrenska University Hospital, Sweden.

出版信息

Phys Med Biol. 1998 Nov;43(11):3183-94. doi: 10.1088/0031-9155/43/11/002.

Abstract

This paper presents two methods for absorbed dose determination with ionization chambers at short distance from 60Co and 192Ir brachytherapy sources. The methods are modifications of the Bragg-Gray and large cavity principles given in the IAEA code of practice for high- and medium-energy photon beams. A non-uniformity correction factor to account for the non-uniform electron fluence in the air cavity is introduced into the methods. The absorbed dose rates were determined from ionization chamber measurements at distances between 1.5 and 5.0 cm from the brachytherapy sources. The agreement between the two methods is excellent in 60Co brachytherapy dosimetry. For 192Ir dosimetry, the difference is less than 2.5% at all distances. In absorbed dose rate calculations with the 60Co brachytherapy source, the ratios between calculated and experimentally determined absorbed dose rates are 0.987 and 0.994 depending on the method used for absorption and scatter correction. In 192Ir dosimetry, the large cavity principle gives almost identical values to those which can be obtained with the AAPM recommendations. Using the chambers according to the Bragg-Gray principle in 192Ir dosimetry, the agreement with AAPM calculated absorbed dose rates is within 2.5% at all distances. The uncertainty, expressed as one standard deviation, in the experimentally determined absorbed dose is estimated to be between 3 and 4%.

摘要

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验