Xie S, Allington R W, Svec F, Fréchet J M
ISCO Inc., Lincoln, NE 68504-1328, USA.
J Chromatogr A. 1999 Dec 31;865(1-2):169-74. doi: 10.1016/s0021-9673(99)00981-4.
Monolithic macroporous poly(styrene-co-divinylbenzene) stationary phases have been prepared by free radical polymerization within the confines of 4.6-mm I.D. chromatographic columns. The optimized porous properties allow the mobile phase to flow through these columns at flow-rates of up to 10 ml/min. As opposed to the simultaneously tested columns packed with either silica or synthetic polymer beads, the monoliths exhibit only modest back pressure. The monolithic columns were able to separate mixtures of peptides and proteins in a very short time. Under the optimized conditions, the separation of five proteins can be easily achieved in less than 20 s.
通过在4.6毫米内径的色谱柱范围内进行自由基聚合反应,制备了整体式大孔聚苯乙烯-二乙烯基苯共聚物固定相。优化后的多孔结构特性使得流动相能够以高达10毫升/分钟的流速通过这些色谱柱。与同时测试的填充硅胶或合成聚合物微球的色谱柱不同,整体柱仅表现出适度的背压。整体柱能够在极短的时间内分离肽和蛋白质的混合物。在优化条件下,能够在不到20秒的时间内轻松实现五种蛋白质的分离。