Kuhn A, Scheller G, Schwarz M
Orthopädische Universitätsklinik Mannheim.
Biomed Tech (Berl). 1999 Dec;44(12):356-9. doi: 10.1515/bmte.1999.44.12.356.
The initial stability of 6 different hemispherical press-fit titanium acetabular cups was investigated by lever-out tests using a foam model. Each cup was implanted in a standardised manner into machined PVC-foam blocks with an underreaming of 1/2/3 mm, and levered out 5 times. A computer recorded the load-displacement curves. Both insertion forces and lever-out forces increased with increasing press-fit from 1 to 2 mm. With underreaming of 2 mm, insertion forces varied between 9 and 34 Nm, and lever-out torques between 13 and 23 Nm. The cup made up of multi-layer irregular titanium wire mesh had the highest initial stability of all the cups tested. 3 mm underreaming was associated with insertion difficulties, so that lever-out is difficult to compare with that with 1 and 2 mm. The amount of contact on the equatorial rim, where we saw the most attrition after lever-out, is the most important factor for determining the lever-out torque. The quality of the PVC foam blocks, the accuracy of reaming, and the definition of cup insertion are very important parameters for test reproduction. Since the trials were conducted under laboratory conditions, translation of the results to the intraoperative situation should be undertaken only with certain reservations.
通过使用泡沫模型的撬出试验,研究了6种不同半球形压配式钛髋臼杯的初始稳定性。每个髋臼杯以标准化方式植入直径扩大1/2/3毫米的加工PVC泡沫块中,并撬出5次。计算机记录载荷-位移曲线。随着压配从1毫米增加到2毫米,插入力和撬出力均增加。当直径扩大2毫米时,插入力在9至34牛米之间变化,撬出扭矩在13至23牛米之间。在所有测试的髋臼杯中,由多层不规则钛丝网制成的髋臼杯具有最高的初始稳定性。直径扩大3毫米会导致插入困难,因此难以与直径扩大1毫米和2毫米时的撬出情况进行比较。赤道边缘的接触量是决定撬出扭矩最重要的因素,在撬出后我们在赤道边缘看到了最多的磨损。PVC泡沫块的质量、扩孔的精度以及髋臼杯插入的定义是测试重复性的非常重要的参数。由于试验是在实验室条件下进行的,因此将结果转化为术中情况时应有所保留。