Lin S J, Chao H C, Yan D C, Huang J L
Department of Pediatrics, Chang Gung Children's Hospital, Taiwan.
Asian Pac J Allergy Immunol. 1999 Dec;17(4):269-73.
We compared the effect of using 2 different serum collecting tubes, serum separation tubes (SST, with clot activator and gel barrier) and conventional glass tubes (with no additives), on circulating levels of eosinophil cationic protein (ECP) in asthmatic children and controls. The serum ECP values obtained from both SST and glass tubes were significantly higher in asthmatic children than in corresponding controls. ECP values were higher in serum samples using SST than in those using glass tubes (P<0.01), while no difference was found between the two in controls. ECP levels correlated with peripheral eosinophil counts, for SST samples and glass tube samples alike. The difference in ECP levels between these two tubes also correlated with circulating eosinophil counts (r = 0.62, P = 0.004) After 18-hour storage at room temperature, the ECP values increased significantly in samples obtained from asthmatic children. No difference in ECP values between SST samples and glass tube samples was found for 18 hour samples. Thus, ECP levels obtained from SST samples and glass tube samples, though reliable, should not be directly compared, especially in asthmatic children with eosinophilia.
我们比较了使用两种不同血清采集管,即血清分离管(SST,含促凝剂和凝胶屏障)和传统玻璃管(无添加剂),对哮喘儿童和对照组嗜酸性粒细胞阳离子蛋白(ECP)循环水平的影响。哮喘儿童从SST管和玻璃管获得的血清ECP值均显著高于相应对照组。使用SST管的血清样本中的ECP值高于使用玻璃管的血清样本(P<0.01),而对照组中两者无差异。SST样本和玻璃管样本的ECP水平均与外周嗜酸性粒细胞计数相关。这两种管之间ECP水平的差异也与循环嗜酸性粒细胞计数相关(r = 0.62,P = 0.004)。在室温下储存18小时后,哮喘儿童样本中的ECP值显著升高。18小时样本的SST样本和玻璃管样本之间的ECP值无差异。因此,从SST样本和玻璃管样本获得的ECP水平虽然可靠,但不应直接比较,尤其是在嗜酸性粒细胞增多的哮喘儿童中。