Bertocchi P, Ciranni E, Compagnone D, Magearu V, Palleschi G, Pirvutoiu S, Valvo L
Laboratorio di Chimica del Farmaco, Istituto Superiore di Sanita, Rome, Italy.
J Pharm Biomed Anal. 1999 Jun;20(1-2):263-9. doi: 10.1016/s0731-7085(99)00032-1.
An enzymatic amperometric procedure for measurement of mercury(II) in pharmaceuticals, based on the inhibition of invertase and on a glucose electrode was studied. Analytical parameters for measurements in batch and flow injection analysis (FIA) have been optimised. Mercury(II) was detected in the 10-60 ppb range with RSD < or =2%. A sample throughput of 6 h(-1) for batch and 15 h(-1) for FIA was obtained. The total mercury(II) from thimerosal (thiomersal, sodium ethylmercurithiosalicylate) in eye-drop samples was measured with the amperometric procedure after oxidative cleavage treatment. Results for both batch and FIA procedures correlated well with atomic absorbtion spectroscopy (AAS) data.