Hepatic MR imaging with a dynamic contrast-enhanced isotropic volumetric interpolated breath-hold examination: feasibility, reproducibility, and technical quality.

作者信息

Lee V S, Lavelle M T, Rofsky N M, Laub G, Thomasson D M, Krinsky G A, Weinreb J C

机构信息

Department of Radiology, New York University Medical Center, 530 First Ave, New York, NY 10016, USA.

出版信息

Radiology. 2000 May;215(2):365-72. doi: 10.1148/radiology.215.2.r00ma16365.

Abstract

PURPOSE

To evaluate the feasibility, reproducibility, and technical quality of a dynamic contrast material-enhanced isotropic three-dimensional (3D) volumetric interpolated breath-hold hepatic magnetic resonance (MR) imaging examination.

MATERIALS AND METHODS

Fifty patients underwent 3D spoiled gradient-echo imaging (4.2/1.8 [repetition time msec/echo time msec]; flip angle, 12 degrees; interpolation in three directions; intermittent fat saturation; pixel size </= 2.5 mm in all dimensions) before and dynamically after administration of gadopentetate dimeglumine, with the first enhanced acquisition timed for hepatic arterial dominance by using a test bolus of contrast material. Qualitative and quantitative measures of image quality were determined. Patterns of arterial and venous anatomy were assessed. Ten patients (20%) underwent repeat imaging within 6 months, and reproducibility was evaluated.

RESULTS

Hepatic contrast-to-noise ratios for nonenhanced and arterial, portal venous, and equilibrium phase studies averaged 13.0 +/- 12.6 (SD), 17.4 +/- 11.8, 30.4 +/- 16.2, and 28.6 +/- 21.1, respectively. During arterial phase, the liver enhanced a mean of 29% of the maximal enhancement as measured during portal venous phase. Hepatic vascular anatomic variants were comparable in distribution to those cited in published articles. Repeat studies were not significantly different in image quality when compared with original studies.

CONCLUSION

High-quality arterial phase 3D volumetric interpolated breath-hold images can be obtained reliably and reproducibly when timed by using a test dose of contrast material.

摘要

目的

评估动态对比剂增强各向同性三维(3D)容积内插屏气肝脏磁共振(MR)成像检查的可行性、可重复性及技术质量。

材料与方法

50例患者在静脉注射钆喷酸葡胺前及注射后进行动态3D扰相梯度回波成像(重复时间4.2/1.8 [毫秒/回波时间毫秒];翻转角12°;三维方向内插;间歇性脂肪抑制;各维度像素大小≤2.5毫米),首次增强采集通过使用对比剂测试团注来确定肝动脉期优势。确定图像质量的定性和定量指标。评估动脉和静脉解剖结构模式。10例患者(20%)在6个月内进行重复成像,并评估可重复性。

结果

非增强、动脉期、门静脉期和平衡期研究的肝脏对比噪声比分别平均为13.0±12.6(标准差)、17.4±11.8、30.4±16.2和28.6±21.1。在动脉期,肝脏增强程度平均为门静脉期测量的最大增强程度的29%。肝脏血管解剖变异的分布与已发表文章中引用的相当。与原始研究相比,重复研究的图像质量无显著差异。

结论

当使用对比剂测试剂量进行定时时,可可靠且可重复地获得高质量的动脉期3D容积内插屏气图像。