Weskott H P
Medizinische Klinik II, Klinikum Hannover.
Ultraschall Med. 2000 Apr;21(2):59-65. doi: 10.1055/s-2000-319.
A new ultrasound-based blood flow detection method, B-Flow is introduced. The amplitude of scatterers in flowing blood is imaged by a subtraction mode of two to four vectors along one line. Noise reduction is gained by using digital encoded US pulses. Signal enhancement is performed by using a pulse compression method (coded excitation). Compared to Color Doppler (CDI) Imaging B-Flow has a higher time (frame rate) and better spatial resolution. Typical CDI artifacts like color blooming or aliasing are missed. B-Flow does not allow velocity measurements. B-Flow has its clinical value in imaging perfused arterial vessels, simultaneous imaging of high and low velocities and delineating plaque morphology with high resolution.
介绍了一种基于超声的新型血流检测方法——B -flow。通过沿一条线的两到四个矢量的减法模式对流动血液中散射体的振幅进行成像。使用数字编码超声脉冲可实现降噪。通过脉冲压缩方法(编码激励)进行信号增强。与彩色多普勒(CDI)成像相比,B-flow 具有更高的时间分辨率(帧率)和更好的空间分辨率。不会出现典型的 CDI 伪像,如彩色外溢或混叠。B-flow 不允许进行速度测量。B-flow 在灌注动脉血管成像、高低速同时成像以及高分辨率描绘斑块形态方面具有临床价值。