Hildestad C A, Morin R L
Department of Radiology, Mayo Clinic Jacksonville, FL 32224, USA.
J Digit Imaging. 2000 May;13(2 Suppl 1):43-4. doi: 10.1007/BF03167622.
Even though facilities using computed radiography (CR) operate in an electronic environment, the production of hard-copy films is still necessary during the transition period, as well as for particular needs following complete implementation. We have implemented a quantitative technique to match the response of printed CR film with that of previous screen/film combinations. A stepwedge is radiographed using the conventional system. The same stepwedge is then radiographed (same geometry and technique) using the CR system. Following processing and printing, the plot of optical density versus step for the CR system is compared with that of the screen/film system. Adjustments are made to the printing parameters until the response curves are identical. All other translation tables in the system are set to be linear. This has proven to be a valuable technique for us and provides CR printed image quality that is equivalent to that of our previous screen/film combinations.
尽管使用计算机X线摄影(CR)的设备在电子环境中运行,但在过渡期间仍需要制作硬拷贝胶片,在完全实施后满足特定需求时也同样如此。我们已经实施了一种定量技术,以使打印的CR胶片的响应与先前的屏/片组合的响应相匹配。使用传统系统对阶梯楔形模进行射线照相。然后使用CR系统对同一阶梯楔形模进行射线照相(相同的几何形状和技术)。处理和打印后,将CR系统的光学密度与阶梯的关系图与屏/片系统的关系图进行比较。调整打印参数,直到响应曲线相同。系统中的所有其他转换表设置为线性。这已被证明对我们是一种有价值的技术,并提供了与我们先前的屏/片组合相当的CR打印图像质量。