Cikajlo I, Bajd T
Faculty of Electrical Engineering, University of Ljubljana, Slovenia.
J Med Eng Technol. 2000 Jan-Feb;24(1):14-9. doi: 10.1080/030919000293996.
A telekinesthetic feedback implemented into functional electrical stimulation (FES) orthosis is described. Single channel FES is used to provoke ankle dorsiflexion during walking. FES is controlled manually by a special lever, built into the handle of the crutch. The angular position of the lever defines the intensity of stimulation and thus the magnitude of the ankle dorsiflexion. The measured joint angle provides the feedback information about the ankle joint position, which is presented to the user as a force feedback applied to the control lever. As the first step in the development of a complex micromechatronic device, a simulated testing environment was prepared. A computer model, comprising dynamic foot characteristics, as well as agonistic and antagonistic muscle groups, substitutes the ankle joint. The model also includes fatiguing of the electrically stimulated muscles. For experimental purposes an actuated control lever was built. The efficacy of the telekinesthetic feedback was evaluated in a group of six healthy persons.
描述了一种应用于功能性电刺激(FES)矫形器的心灵感应反馈。单通道FES用于在行走过程中引发踝关节背屈。FES由内置在拐杖手柄中的特殊杠杆手动控制。杠杆的角位置定义了刺激强度,从而决定了踝关节背屈的幅度。测量的关节角度提供有关踝关节位置的反馈信息,该信息作为施加在控制杠杆上的力反馈呈现给用户。作为开发复杂微机电装置的第一步,准备了一个模拟测试环境。一个包含动态足部特征以及主动和拮抗肌群的计算机模型替代了踝关节。该模型还包括电刺激肌肉的疲劳。为了实验目的,制作了一个驱动控制杠杆。在一组六名健康人身上评估了心灵感应反馈的效果。