Ohto T, Yamamoto F, Nakajima N
First Department of Surgery, Chiba University School of Medicine, Japan.
Jpn J Thorac Cardiovasc Surg. 2000 May;48(5):295-300. doi: 10.1007/BF03218142.
To prevent postoperative pulmonary dysfunction, we have investigated the effect of the Leuko-Guard 6 leukocyte-reducing arterial line filter (LG6) on postoperative lung function.
Twenty-six cases of adult valvular heart disease were included in this study. Thirteen cases were operated upon using the LG6 (Group LG), and 13 cases were operated upon using a conventional arterial line filter (Group C). Neutrophil, polymorphonuclear leukocyte elastase and lipoperoxide were measured for this study, and the lung function was evaluated using the Oxygenation Index (PaO2/FiO2).
Statistically significant differences were observed in neutrophil counts between Group LG and Group C (LG = 2225 +/- 572/mm3, C = 3157 +/- 1413/mm3, p = 0.04) at 5 minutes after the onset of cardiopulmonary bypass. In simultaneous blood sampling from the pulmonary artery and the pulmonary vein, the sequestration of neutrophil in the lung decreased in Group LG after the discontinuation of cardiopulmonary bypass. Release of polymorphonuclear leukocyte elastase from the lungs was significantly decreased (p = 0.04) in the Group LG at 1 hour post-bypass. Significant differences were observed in the Oxygenation Index between Group LG and Group C (LG = 398 +/- 72, C = 326 +/- 71, p = 0.019) at 3 hours post-bypass.
We concluded that LG6 improved the postoperative lung function, and its mechanism might be derived from the prevention of leukosequestration in the lungs that occurs during the rewarming phase due to selective absorption of activated leukocyte by the LG6.
为预防术后肺功能障碍,我们研究了Leuko - Guard 6白细胞滤除动脉管路过滤器(LG6)对术后肺功能的影响。
本研究纳入26例成年瓣膜性心脏病患者。13例患者手术中使用LG6(LG组),13例患者手术中使用传统动脉管路过滤器(C组)。本研究检测了中性粒细胞、多形核白细胞弹性蛋白酶和脂过氧化物,并使用氧合指数(PaO2/FiO2)评估肺功能。
体外循环开始5分钟时,LG组和C组中性粒细胞计数存在统计学显著差异(LG = 2225 ± 572/mm3,C = 3157 ± 1413/mm3,p = 0.04)。在同时从肺动脉和肺静脉采集血样时,体外循环停止后LG组肺内中性粒细胞的扣押减少。体外循环后1小时,LG组肺内多形核白细胞弹性蛋白酶的释放显著减少(p = 0.04)。体外循环后3小时,LG组和C组的氧合指数存在显著差异(LG = 398 ± 72,C = 326 ± 71,p = 0.019)。
我们得出结论,LG6改善了术后肺功能,其机制可能源于LG6对活化白细胞的选择性吸附,从而预防了复温阶段肺内白细胞扣押的发生。