Suppr超能文献

Transdermal extraction of analytes using low-frequency ultrasound.

作者信息

Mitragotri S, Coleman M, Kost J, Langer R

机构信息

Department of Chemical Engineering, University of California, Santa Barbara 93106, USA.

出版信息

Pharm Res. 2000 Apr;17(4):466-70. doi: 10.1023/a:1007537222591.

Abstract

PURPOSE

Transdermal extraction of clinically relevant analytes offers a potentially non-invasive method of diagnostics. However, development of such a method is limited by the low skin permeability. In this paper, we report a potential method for non-invasive diagnostics based on ultrasonic skin permeabilization and subsequent extraction of interstitial fluid (ISF) across the skin.

METHODS

In vivo experiments were performed using Sprague Dawley rats to assess ultrasound-induced skin permeabilization and subsequent extraction of various analytes. Serum and ISF concentrations of various analytes were measured.

RESULTS

Application of low-frequency ultrasound rapidly increased skin permeability. Skin remained in a state of high permeability for at least three hours. During this period, application of vacuum extracted ISF across rat skin in vivo at a rate of 25.7 microl/cm2/hr. We measured concentrations of various analytes including glucose, albumin, calcium, urea, triglycerides, lactate, and dextran in transdermally extracted fluid. The composition of the fluid extracted transdermally is similar to that of ISF.

CONCLUSIONS

Application of low-frequency ultrasound allows skin permeabilization and extraction of ISF across the skin.