Parkes J L, Slatin S L, Pardo S, Ginsberg B H
Becton Dickinson and Company, Franklin Lakes, New Jersey 07417, USA.
Diabetes Care. 2000 Aug;23(8):1143-8. doi: 10.2337/diacare.23.8.1143.
The objectives of this study were 1) to construct new error grids (EGs) for blood glucose (BG) self-monitoring by using the expertise of a large panel of clinicians and 2) to use the new EGs to evaluate the accuracy of BG measurements made by patients.
To construct new EGs for type 1 and type 2 diabetic patients, a total of 100 experts of diabetes were asked to assign any error in BG measurement to 1 of 5 risk categories. We used these EGs to evaluate the accuracy of self-monitoring of blood glucose (SMBG) levels in 152 diabetic patients. The SMBG data were used to compare the new type 1 diabetes EG with a traditional EG.
Both the type 1 and type 2 diabetes EGs divide the risk plane into 8 concentric zones with no discontinuities. The new EGs are similar to each other, but they differ from the traditional EG in several significant ways. When used to evaluate a data set of measurements made by a sample of patients experienced in SMBG, the new type 1 diabetes EG rated 98.6% of their measurements as clinically acceptable, compared with 95% for the traditional EG.
The consensus EGs furnish a new tool for evaluating errors in the measurement of BG for patients with type 1 and type 2 diabetes.
本研究的目的是1)利用大量临床医生的专业知识构建用于血糖(BG)自我监测的新误差网格(EGs),以及2)使用新的EGs评估患者进行的BG测量的准确性。
为1型和2型糖尿病患者构建新的EGs,共邀请了100位糖尿病专家将BG测量中的任何误差归入5个风险类别中的1个。我们使用这些EGs评估了152例糖尿病患者自我血糖监测(SMBG)水平的准确性。利用SMBG数据将新型1型糖尿病EG与传统EG进行比较。
1型和2型糖尿病EGs均将风险平面划分为8个同心区域,无间断。新的EGs彼此相似,但在几个重要方面与传统EG不同。当用于评估一组由有SMBG经验的患者样本进行的测量数据时,新型1型糖尿病EG将其98.6%的测量结果评为临床可接受,而传统EG为95%。
共识EGs为评估1型和2型糖尿病患者BG测量误差提供了一种新工具。