作为多毛细管气相色谱柱与飞行时间质谱仪之间接口的喷射分离器和开放式分流器的比较。

Comparison of a jet separator and an open splitter as an interface between a multi-capillary gas chromatographic column and a time-of-flight mass spectrometer.

作者信息

Pongpun N, Mlynski V, Crisp PT, Guilhaus M

机构信息

School of Chemistry, University of New South Wales, Sydney 2052, Australia.

出版信息

J Mass Spectrom. 2000 Sep;35(9):1105-11. doi: 10.1002/1096-9888(200009)35:9<1105::AID-JMS35>3.0.CO;2-W.

Abstract

A gas chromatographic/time-of-flight mass spectrometric (GC/TOFMS) interface is being developed for fast on-line analysis utilizing multi-capillary column technology. A variable gap-distance jet separator has been constructed and its performance compared with that of a commercially supplied post-column open splitter recommended for use between the multi-capillary column and a mass spectrometer. Both interfaces were found to be compatible with the GC/TOFMS system at high carrier gas flow-rates, facilitating high-speed and high-resolution separations. The systems were investigated and tested with a mixture of volatile organic compounds (VOCs) with molecular masses from 85 to 166: dichloromethane, toluene, m-dichlorobenzene, o-dichlorobenzene and tetrachloroethylene. The optimum tip-to-tip gap distance corresponding to the highest efficiency of the jet separator was found to be 0.030 mm for each compound at carrier gas flow-rates of 20, 40 and 60 ml min(-1) giving, in the ion source housing, ion gauge pressure readings of 1.6 x 10(-6), 5.0 x 10(-6) and 5.8 x 10(-6) mbar, respectively. The efficiency of the jet separator (10-30% yields) was significantly higher than that of the open splitter (6-9% yields). The observation that the open splitter did not provide a constant flow-rate to the ion source was not in agreement with the manufacturer's specifications. A method for measuring the gas flow-rates in all parts of the equipment is described. The correlation between yield in the jet separator and molecular mass for the heterogeneous set of compounds studied was found to be less linear than usually reported for homologous series of compounds in jet separator studies. The result suggests that the pressure conditions in the jet may be sufficient for the separation process to be partly controlled by diffusion rather than predominately by effusion. Copyright 2000 John Wiley & Sons, Ltd.

摘要

正在开发一种气相色谱/飞行时间质谱联用(GC/TOFMS)接口,用于利用多毛细管柱技术进行快速在线分析。已构建了一种可变间隙距离的喷射分离器,并将其性能与市售的柱后开放式分流器进行了比较,该分流器推荐用于多毛细管柱和质谱仪之间。发现这两种接口在高载气流速下均与GC/TOFMS系统兼容,便于进行高速和高分辨率分离。使用分子量从85到166的挥发性有机化合物(VOCs)混合物对系统进行了研究和测试:二氯甲烷、甲苯、间二氯苯、邻二氯苯和四氯乙烯。发现在20、40和60 ml min(-1)的载气流速下,喷射分离器效率最高时对应的最佳尖端到尖端间隙距离对于每种化合物均为0.030 mm,在离子源外壳中,离子规压力读数分别为1.6 x 10(-6)、5.0 x 10(-6)和5.8 x 10(-6) mbar。喷射分离器的效率(产率为10 - 30%)明显高于开放式分流器(产率为6 - 9%)。观察到开放式分流器未向离子源提供恒定流速,这与制造商的规格不一致。描述了一种测量设备各部分气体流速的方法。在所研究的一组非均相化合物中,发现喷射分离器中的产率与分子量之间的相关性不如喷射分离器研究中通常报道的同系物系列化合物那样呈线性。结果表明,喷射器中的压力条件可能足以使分离过程部分受扩散控制而非主要受泻流控制。版权所有2000 John Wiley & Sons, Ltd.

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验