Hayashi K, Hayashi H, Nakao F, Hayashi F
Hayashi Eye Hospital, Fukuoka, Japan.
Am J Ophthalmol. 2000 Oct;130(4):477-82. doi: 10.1016/s0002-9394(00)00526-2.
To examine the influence of astigmatism on the visual acuity of patients with multifocal and monofocal intraocular lenses.
Thirty eyes of 30 patients who underwent five-zone refractive multifocal intraocular lens implantation and 30 eyes of 30 age-matched patients who had monofocal intraocular lens implantation were included. The visual acuities of these patients at 5.0, 3.0, 2.0, 1.0, 0.7, 0.5, and 0.3 m were measured by means of an all-distance vision tester after addition of a cylindrical lens of 0, 0.5, 1.0, 1.5, 2.0, and 2.5 diopters.
Mean visual acuity at all distances decreased in proportion to the diopters of astigmatism in both the multifocal and monofocal intraocular lens groups. When astigmatism was 0.5, 1.0, or 1.5 diopters, distance visual acuity in the multifocal group was significantly worse than that in the monofocal group; with astigmatism of 2.0 diopters or more, intermediate visual acuity was also worse in the multifocal group. In contrast, near visual acuity in the multifocal group was significantly better than that in the monofocal group at all astigmatic diopters. When astigmatism was within 1.0 diopter, visual acuity in the multifocal group reached 20/29 at distance and 20/50 at near. Mean contrast sensitivity was also worse in the multifocal group than in the monofocal group.
Both distance and intermediate visual acuity deterioration caused by astigmatism was greater with a multifocal intraocular lens than with a monofocal intraocular lens, whereas near visual acuity was better with the multifocal intraocular lens. When astigmatism was within 1.0 diopter, eyes with a multifocal intraocular lens achieved good visual acuity at both distance and near.
研究散光对多焦点和单焦点人工晶状体植入患者视力的影响。
纳入30例行五区屈光性多焦点人工晶状体植入术患者的30只眼,以及30例年龄匹配的行单焦点人工晶状体植入术患者的30只眼。在添加0、0.5、1.0、1.5、2.0和2.5屈光度的柱面透镜后,使用全距离视力测试仪测量这些患者在5.0、3.0、2.0、1.0、0.7、0.5和0.3米处的视力。
多焦点和单焦点人工晶状体组中,所有距离的平均视力均随散光屈光度的增加而下降。当散光为0.5、1.0或1.5屈光度时,多焦点组的远视力明显低于单焦点组;散光为2.0屈光度及以上时,多焦点组的中视力也较差。相比之下,在所有散光屈光度下,多焦点组的近视力均明显优于单焦点组。当散光在1.0屈光度以内时,多焦点组的远视力达到20/29,近视力达到20/50。多焦点组的平均对比敏感度也低于单焦点组。
散光导致的多焦点人工晶状体组远视力和中视力下降程度大于单焦点人工晶状体组,而多焦点人工晶状体组的近视力较好。当散光在1.0屈光度以内时,多焦点人工晶状体眼的远近视力均良好。