Assisted venous drainage presents the risk of undetected air microembolism.

作者信息

Lapietra A, Grossi E A, Pua B B, Esposito R A, Galloway A C, Derivaux C C, Glassman L R, Culliford A T, Ribakove G H, Colvin S B

机构信息

Department of Surgery, Division of Cardiothoracic Surgery, New York University School of Medicine, New York, NY, USA.

出版信息

J Thorac Cardiovasc Surg. 2000 Nov;120(5):856-62. doi: 10.1067/mtc.2000.110183.

Abstract

OBJECTIVES

The proliferation of minimally invasive cardiac surgery has increased dependence on augmented venous return techniques for cardiopulmonary bypass. Such augmented techniques have the potential to introduce venous air emboli, which can pass to the patient. We examined the potential for the transmission of air emboli with different augmented venous return techniques.

METHODS

In vitro bypass systems with augmented venous drainage were created with either kinetically augmented or vacuum-augmented venous return. Roller or centrifugal pumps were used for arterial perfusion in combination with a hollow fiber oxygenator and a 40-micrometer arterial filter. Air was introduced into the venous line via an open 25-gauge needle. Test conditions involved varying the amount of negative venous pressure, the augmented venous return technique, and the arterial pump type. Measurements were recorded at the following sites: pre-arterial pump, post-arterial pump, post-oxygenator, and patient side.

RESULTS

Kinetically augmented venous return quickly filled the centrifugal venous pump with macrobubbles requiring continuous manual clearing; a steady state to test for air embolism could not be achieved. Vacuum-augmented venous return handled the air leakage satisfactorily and microbubbles per minute were measured. Higher vacuum pressures resulted in delivery of significantly more microbubbles to the "patient" (P <.001). The use of an arterial centrifugal pump was associated with fewer microbubbles (P =.02).

CONCLUSIONS

Some augmented venous return configurations permit a significant quantity of microbubbles to reach the patient despite filtration. A centrifugal pump has air-handling disadvantages when used for kinetic venous drainage, but when used as an arterial pump in combination with vacuum-assisted venous drainage it aids in clearing air emboli.