Jorhem L, Engman J
National Food Administration, Uppsala, Sweden.
J AOAC Int. 2000 Sep-Oct;83(5):1189-203.
A method for determination of lead, cadmium, zinc, copper, and iron by atomic absorption spectrometry (AAS) after microwave digestion was subjected to a collaborative study in which 16 laboratories participated [including users of inductively coupled plasma (ICP) and ICP-mass spectrometry (MS)]. The types of samples included in the study were minced fish, wheat bran, milk powder, bovine and pig liver, mushroom, 2 simulated diets, and bovine muscle; the last 4 were certified reference materials. These were analyzed as single (4 samples), double blind (1 sample), or split level (2 samples) samples. Before the collaborative study, a pretrial was conducted in which 4 ready-made solutions and one fish tissue sample were analyzed for Pb and Cu. The reproducibility relative standard deviation (RSDR) values, for results above the detection limit, ranged from 59% at 0.155 mg/kg to 16% at 1.62 mg/kg for Pb, from 28% at 0.0124 mg/kg to 11% at 0.482 mg/kg for Cd, from 9.3% at 35.3 mg/kg to 1.7% at 147 mg/kg for Zn, from 39% at 0.241 mg/kg to 3.0% at 63.4 mg/kg for Cu, and from 17% at 7.4 mg/kg to 5.9% at 303 mg/kg for Fe. The RSDR values agreed well with the norms described by the International Union of Pure and Applied Chemistry. As a complement to the AAS determinations, a number of laboratories analyzed the samples either by ICP or by ICP-MS. The results of these analyses agreed well with the AAS results. On the basis of the results of the collaborative study, the method was adopted Official First Action by AOAC INTERNATIONAL.
一种通过微波消解后用原子吸收光谱法(AAS)测定铅、镉、锌、铜和铁的方法进行了一项协作研究,有16个实验室参与(包括电感耦合等离子体(ICP)和ICP-质谱(MS)的用户)。研究中包括的样品类型有碎鱼、麦麸、奶粉、牛和猪肝、蘑菇、2种模拟日粮以及牛肌肉;最后4种是有证标准物质。这些样品作为单样(4个样品)、双盲样(1个样品)或分割水平样(2个样品)进行分析。在协作研究之前,进行了一次预试验,分析了4种现成溶液和1个鱼组织样品中的铅和铜。对于高于检测限的结果,重复性相对标准偏差(RSDR)值,铅在0.155 mg/kg时为59%,在1.62 mg/kg时为16%;镉在0.0124 mg/kg时为28%,在0.482 mg/kg时为11%;锌在35.3 mg/kg时为9.3%,在147 mg/kg时为1.7%;铜在0.241 mg/kg时为39%,在63.4 mg/kg时为3.0%;铁在7.4 mg/kg时为17%,在303 mg/kg时为5.9%。RSDR值与国际纯粹与应用化学联合会描述的规范吻合良好。作为AAS测定的补充,一些实验室通过ICP或ICP-MS对样品进行了分析。这些分析结果与AAS结果吻合良好。根据协作研究结果,该方法被美国官方分析化学家协会(AOAC INTERNATIONAL)采用为官方首次行动方法。